      "status": "Ongoing",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91849",
      "recalling_firm": "Stryker Corporation",
      "address_1": "1941 Stryker Way",
      "address_2": "",
      "postal_code": "49002-9711",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1341-2023",
      "product_description": "Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 34X4,1BLA,1PRT QUICK  Catalog number: 5921-034-135NS    Indicated for use in surgical procedures that require the temporary occlusion of blood flow in a patient s extremities",
      "product_quantity": "1979 units (packs of 10)",
      "reason_for_recall": "Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications",
      "recall_initiation_date": "20230301",
      "center_classification_date": "20230404",
      "report_date": "20230412",
      "code_info": "GTIN:  07613327606171 Lot Numbers: 2022111805 2022112801 2022112803 2022112804 2022112805 2022112806 2022112807 2022112808 2022120506 2022120507 2022120508",
      "more_code_info": ""
    }
  ]
}