  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Twinsburg",
      "address_1": "1959 Summit Commerce Park",
      "reason_for_recall": "These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.",
      "address_2": "",
      "product_quantity": "18",
      "code_info": "Serial Number: G3007689, G3007688, G3007687, 206H2628, 206H2629,  206H2630, 206H2620, 206H2621, 206H2622, 206H2623, 206H26224,  206H2625, 206H2626, 206H2627, 205X3897, 205X3898, 205X3899, and  205X3900",
      "center_classification_date": "20200220",
      "distribution_pattern": "Domestic distribution to CA, FL, GA, MA, NV, OH, TX, and WV. Foreign distribution to Canada.",
      "state": "OH",
      "product_description": "UST-2265-2 Probe, Part Number: UST-2265-2 - Product Usage: The UST-2265-2 probe is intended to be used for diagnostic ultrasound imaging or fluid flow analysis of the human body as Continuous-wave Doppler (CWD) for Cardiac Adult, Cardiac Pediatric, and Peripheral Vessel.",
      "report_date": "20200226",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hitachi Medical Systems America Inc",
      "recall_number": "Z-1341-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "84838",
      "termination_date": "20200731",
      "more_code_info": "",
      "recall_initiation_date": "20200131",
      "postal_code": "44087-2371",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}