      "country": "United States",
      "city": "Malvern",
      "address_1": "40 Liberty Blvd",
      "reason_for_recall": "Potential exists for one or more axes of the robotic stand to become blocked by their safety brakes;  In the event the axis is blocked by its safety brake, the robotic stand can no longer perform the intended  movements",
      "address_2": "",
      "product_quantity": "96 US",
      "code_info": "All serial numbers",
      "center_classification_date": "20200220",
      "distribution_pattern": "US Nationwide distribution.",
      "state": "PA",
      "product_description": "ARTIS Pheno-Interventional Fluoroscopic X-Ray System  Model Number: 10849000 - Product Usage:AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Axiom Artis can also support the acquisition of position triggered imaging for spatial data synthesis.",
      "report_date": "20200226",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-1340-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "84858",
      "termination_date": "20200710",
      "more_code_info": "",
      "recall_initiation_date": "20200124",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}