  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Ambler",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91782",
      "recalling_firm": "Moberg Research, Inc.",
      "address_1": "224 S Maple Way",
      "address_2": "",
      "postal_code": "19002-5528",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of France, Germany, Italy, Sweden.",
      "recall_number": "Z-1338-2023",
      "product_description": "Moberg Component Neuromonitoring System (CNS Monitor)-Intended as a short-term neurological monitoring system that collects, displays, and stores multiple physiological measurements  Model: CNS-350",
      "product_quantity": "89 units",
      "reason_for_recall": "Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin.  DOC",
      "recall_initiation_date": "20230220",
      "center_classification_date": "20230404",
      "report_date": "20230412",
      "code_info": "UDI Number:  B485CNS3501 Serial Numbers: A-CNS-22502   A-CNS-22506   A-CNS-22507   A-CNS-22508   A-CNS-22509   A-CNS-22510   A-CNS-22511   A-CNS-22512   A-CNS-22513   A-CNS-22525   A-CNS-22526   A-CNS-22527   A-CNS-22528   A-CNS-22529   A-CNS-22530   A-CNS-22534   A-CNS-22535   A-CNS-22536   A-CNS-22537   A-CNS-22541   A-CNS-22542   A-CNS-22543   A-CNS-22544      A-CNS-22545   A-CNS-22546   A-CNS-22547   A-CNS-22548   A-CNS-22549   A-CNS-22550   A-CNS-22551   A-CNS-22552   A-CNS-22553   A-CNS-22554   A-CNS-22555   A-CNS-22556   A-CNS-22557   A-CNS-22558   A-CNS-22559   A-CNS-22560   A-CNS-22561   A-CNS-22562   A-CNS-22564   A-CNS-22565   A-CNS-22566   A-CNS-22567        A-CNS-22568    A-CNS-22569    A-CNS-22570    A-CNS-22571    A-CNS-22572    A-CNS-22573    A-CNS-22574    A-CNS-22575    A-CNS-22577    A-CNS-22578    A-CNS-22584    A-CNS-22585    A-CNS-22586    A-CNS-22587    A-CNS-22588    A-CNS-22589    A-CNS-22590    A-CNS-22591",
      "more_code_info": ""
    }
  ]
}