  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67758",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA and the countries of Australia, Belgium, Brazil, Chile, China, Columbia, Denmark,  France, Germany, India, Italy, Lebanon,  Monaco, Netherlands, Norway, Romania, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom.",
      "recall_number": "Z-1334-2014",
      "product_description": "5X250 MM Curved Cannula, Arm 1  for use with daVinci SI Surgical System, model number IS3000.    Intuitive Surgical  1266 Kifer Road  Sunnyvale, CA 94086    The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation,clip- ligation, electrocautery and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators,monopolar cautery, bipolar cautery, 5mm curved cannulae,5mm and 10 mm straight cannulae, flexible blunt obturators, and the 5mm Single-Site Port.",
      "product_quantity": "5217 total cannulae",
      "reason_for_recall": "Intuitive Surgical has become aware of the potential for certain lots of the Single-Site 5mm Curved Cannulae to become damaged during use. The second part of this action is to update Single-Site labeling. These labeling changes instruct users how to inspect cannulae prior to use and to appropriately handle cannulae during use.",
      "recall_initiation_date": "20140313",
      "center_classification_date": "20140404",
      "termination_date": "20140903",
      "report_date": "20140416",
      "code_info": "Part number 428071-03:  Lot numbers:   VE113001  VE121004  VE121205  VE122004  VE122604  VE123014  VE113401  VE121007  VE121407  VE122005  VE122901  VE123107  VE114001  VE121103  VE121410  VE122104  VE122904  VE123202  VE114401  VE121105  VE121604  VE122204  VE122908  VE120602  VE121109  VE121608  VE122302  VE122911  VE120701  VE121110  VE121804  VE122403  VE123007."
    }
  ]
}