  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67758",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA and the countries of Australia, Belgium, Brazil, Chile, China, Columbia, Denmark,  France, Germany, India, Italy, Lebanon,  Monaco, Netherlands, Norway, Romania, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom.",
      "recall_number": "Z-1333-2014",
      "product_description": "5X300 MM Curved Cannula, Arm 2  for use with daVinci SI Surgical System, model number IS3000.    Intuitive Surgical  1266 Kifer Road  Sunnyvale, CA 94086    The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation,clip- ligation, electrocautery and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators,monopolar cautery, bipolar cautery, 5mm curved cannulae,5mm and 10 mm straight cannulae, flexible blunt obturators, and the 5mm Single-Site Port.",
      "product_quantity": "5217 total cannulae",
      "reason_for_recall": "Intuitive Surgical has become aware of the potential for certain lots of the Single-Site 5mm Curved Cannulae to become damaged during use. The second part of this action is to update Single-Site labeling. These labeling changes instruct users how to inspect cannulae prior to use and to appropriately handle cannulae during use.",
      "recall_initiation_date": "20140313",
      "center_classification_date": "20140404",
      "termination_date": "20140903",
      "report_date": "20140416",
      "code_info": "Part number 428062-03:  Lot numbers:   VE105003  VE120304  VE121402  VE121903  VE122704  VE123101  VE111202  VE120402  VE121502  VE121905  VE122705  VE123103  VE114502  VE120504  VE121506  VE122201  VE122707  VE123105  VE114602  VE120902  VE121512  VE122202  VE122903  VE115202  VE121003  VE121606  VE122303  VE123005  VE120202  VE121113  VE121702  VE122501  VE123012."
    }
  ]
}