{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Centennial",
      "address_1": "6248 S Troy Cir Ste A",
      "reason_for_recall": "The pouch integrity of the sterile product can't be assured without a visual examination due to possibly containing a cut or hole in the pouch.",
      "address_2": "",
      "product_quantity": "6,553 units were distributed for all catalog numbers",
      "code_info": "Lot numbers 61508, 61636, 61683, 61753, 61804, 61812, 61813, 61814, 61815, 61862, 61917, 61949, 61991, 62052, 62089, 62134, and 62156",
      "center_classification_date": "20190515",
      "distribution_pattern": "Distribution was nationwide, including PR.   There was government/military distribution.    Foreign distribution was made to Australia, Austria, Belgium, Denmark, Germany, Greece, India, Italy, Japan, Netherlands, New Zealand, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, and United Kingdom,",
      "state": "CO",
      "product_description": "Tissu-Trans FILTRON 500, Catalog 3-TT-FILTRON 500, sterile, disposable, one-patient use.  The firm name on the label is Shippert Medical Technologies, Centennial, CO.",
      "report_date": "20190522",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Shippert Medical Technologies",
      "recall_number": "Z-1332-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81837",
      "more_code_info": "",
      "recall_initiation_date": "20180202",
      "postal_code": "80111-6485",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Completed"
    }
  ]
}