  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67758",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA and the countries of Australia, Belgium, Brazil, Chile, China, Columbia, Denmark,  France, Germany, India, Italy, Lebanon,  Monaco, Netherlands, Norway, Romania, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom.",
      "recall_number": "Z-1332-2014",
      "product_description": "5X300 MM Curved Cannula, Arm 1  for use with daVinci SI Surgical System, model number IS3000.    Intuitive Surgical  1266 Kifer Road  Sunnyvale, CA 94086    The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation,clip- ligation, electrocautery and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators,monopolar cautery, bipolar cautery, 5mm curved cannulae,5mm and 10 mm straight cannulae, flexible blunt obturators, and the 5mm Single-Site Port.",
      "product_quantity": "5217 total cannulae",
      "reason_for_recall": "Intuitive Surgical has become aware of the potential for certain lots of the Single-Site 5mm Curved Cannulae to become damaged during use. The second part of this action is to update Single-Site labeling. These labeling changes instruct users how to inspect cannulae prior to use and to appropriately handle cannulae during use.",
      "recall_initiation_date": "20140313",
      "center_classification_date": "20140404",
      "termination_date": "20140903",
      "report_date": "20140416",
      "code_info": "Part number 428061-03:  Lot numbers: VE105001  VE120201  VE121301  VE121803  VE122502  VE123106  VE105002  VE120301  VE121304  VE121907  VE122605  VE111201  VE120503  VE121403  VE122002  VE122606  VE114501  VE120703  VE121508  VE122101  VE122902  VE114601  VE121002  VE121605  VE122203  VE123006  VE115201  VE121112  VE121701  VE122401  VE123104."
    }
  ]
}