  "results": [
    {
      "status": "Terminated",
      "city": "Mountain View",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67741",
      "recalling_firm": "Siemens Medical Solutions USA, Inc.",
      "address_1": "685 E Middlefield Rd",
      "address_2": "N/A",
      "postal_code": "94043-4045",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Class II Recall - Worldwide Distribution - US Distribution and the countries of Australia, Austria, Belgium, Bhutan, Bolivia, Botswana, Bulgaria, Canada, Chile, Colombia, Croatia, Dem. Rep. Congo, Denmark, Ecuador, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Ivory Coast, Japan, Kuwait, Malaysia, Mauritius, Mongolia, Morocco, Mozambique, Myanmar, Namibia, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Republic Korea, Romania, Saudi Arabia, Serbia, Seychelles, Singapore, South Africa, Spain, Switzerland, Taiwan, Trinidad,Tobago, Turkey, U.A.E. and United Kingdom.",
      "recall_number": "Z-1331-2014",
      "product_description": "Siemens Medical Solutions,  ACUSON P300 Ultrasound system with P300 PA 230E cardiac probe (Esaote product 7348).       Provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.",
      "product_quantity": "675",
      "reason_for_recall": "ACUSON P300 ultrasound systems using the PA230 transducer with a user-defined preset which includes the 100% setting for output power could cause the transducer surface temperature to exceed the maximum allowable limit.",
      "recall_initiation_date": "20140307",
      "center_classification_date": "20140403",
      "termination_date": "20151112",
      "report_date": "20140409",
      "code_info": "Model number 10433816 - ACUSON P3000 Ultrasound System;  Model numbers 10852384 - P300 PA 320E transducer.  Serial numbers:  105  106  114  116  117  118  119  120  121  122  123  124  125  126  127  128  129  130  131  132  133  134  135  136  137  138  139  140  141  142  143  144  145  146  148  149  150  152  153  154  155  156  157  158  159  160  161  162  163  164  165  166  167  168  169  170  171  172  173  174  175  176  178  186  187  188  189  193  194  197  198  199  200  201  202  203  204  205  206  207  208  209  210  211  212  213  214  215  216  217  218  219  220  221  222  223  224  225  226  227  229  230  231  232  233  234  235  236  237  238  239  240  241  242  243  244  245  246  247  248  249  250  251  252  253  254  257  258  259  260  262  263  264  265  266  267  268  269  270  271  272  273  274  275  276  277  278  279  280  281  282  283  284  285  286  287  288  289  290  291  292  293  294  295  296  297  298  299  300  301  302  303  304  305  306  307  308  309  310  311  312  313  314  315  316  317  318  319  320  321  322  323  324  325  330  331  334  336  337  339  340  341  342  343  344  345  347  348  349  350  351  352  353  354  355  356  357  358  359  360  361  362  363  364  365  366  367  368  369  371  373  374  375  377  378  379  380  382  383  384  385  386  387  388  389  390  391  392  393  394  395  396  400  401  402  403  404  405  406  407  408  409  410  411  412  413  414  415  416  417  418  421  422  423  424  425  426  427  428  429  430  431  432  433  434  435  436  437  438  439  440  441  442  443  444  445  446  447  448  449  450  451  452  453  454  455  456  457  458  459  460  461  462  463  464  465  466  467  468  469  470  471  472  475  476  477  478  479  480  481  482  483  484  485  486  487  488  489  490  491  492  493  494  495  496  497  498  501  506  507  508  509  510  511  512  513  515  516  518  519  520  521  522  526  527  529  530  535  536  537  538  539  540  541  542  544  545  546  547  548  549  550  551  552  553  554  555  556  557  558  559  560  561  562  563  564  565  566  567  568  569  596  597  598  601  602  603  604  605  606  607  608  610  614  615  616  617  618  619  620  621  622  624  627  628  629  630  633  634  635  636  637  638  639  640  641  642  643  644  646  647  648  650  651  652  653  654  655  656  657  659  660  661  662  663  664  665  666  667  668  669  670  671  672  674  675  676  678  679  680  681  682  683  684  685  686  687  689  693  694  696  697  698  699  704  706  707  709  710  711  712  713  718  719  720  721  722  723  724  725  727  728  729  730  731  732  733  734  737  739  743  745  746  747  748  749  750  751  752  753  754  755  757  758  759  760  761  762  763  764  765  766  767  768  769  770  771  772  773  775  776  777  778  779  782  783  784  785  791  792  793  794  795  796  797  798  800  801  806  807  808  810  811  812  814  815  816  817  818  819  821  822  823  824  825  826  827  828  829  830  832  834  835  836  837  838  839  840  841  842  843  845  847  848  853  854  855  856  857  858  860  861  862  863  864  865  866  872  879  881  890  892  893  894  895  896  897  898  900  901  902  903  904  909  910  911  912  913  914  919  920  929  933  934  935  936  937  938  939  940  941  942  943  944  945  946  947  948  949  950  951  952  957  958  959  960  961  962  963  964  965  967  969  970  972  973  974  975  976  977  979  980  981  985  986."
    }
  ]
}