  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Glendora",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67849",
      "recalling_firm": "Ormco/Sybronendo",
      "address_1": "1332 S Lone Hill Ave",
      "address_2": "N/A",
      "postal_code": "91740-5339",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Internationally Distributed in the country of Japan.",
      "recall_number": "Z-1330-2014",
      "product_description": "Spirit MB Bracket, UR Central +14T +5A.018, Part Number 494-0110, Lot Number 081351013.    Product Usage:  An orthodontic bracket. Spirit MB Brackets are a fixed attachment on a tooth which holds an archwire during orthodontic treatment.",
      "product_quantity": "150 units",
      "reason_for_recall": "Ormco Corporation initiated this recall of Spirit MB Brackets, Part Number 494-0110, Lot Number 081351013, because the ID dot, which designates the bracket as distal/gingival, is misplaced. The misidentification may lead an orthodontist to place the bracket on the wrong tooth, leading to movement 10 degrees in the wrong direction.",
      "recall_initiation_date": "20140325",
      "center_classification_date": "20140403",
      "termination_date": "20140418",
      "report_date": "20140409",
      "code_info": "Part Number 494-0110, Lot Number 081351013."
    }
  ]
}