  "results": [
    {
      "status": "Ongoing",
      "city": "Hamburg",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84773",
      "recalling_firm": "Philips Medical Systems Gmbh, DMC",
      "address_1": "Development And Manufacturing Ctr.",
      "address_2": "Rontgenstr. 24",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution \u0013 US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MI, MN, MO, MT, NC, NJ, NY, OH, TX, UT, VA and countries of Canada, Australia, Austria, Bangladesh, Belgium, Bulgaria, Chile, China, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, Egypt, Ethiopia, France, Germany, Ghana, Hungary, Iran, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya,   Luxembourg, Malaysia, Martinique, Mexico, Nepal, Netherlands, New Zealand, Nigeria, Oman, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Viet Nam.",
      "recall_number": "Z-1329-2020",
      "product_description": "CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.",
      "product_quantity": "131",
      "reason_for_recall": "When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.",
      "recall_initiation_date": "20191219",
      "center_classification_date": "20200220",
      "termination_date": "20200618",
      "report_date": "20200226",
      "code_info": "(domestic only)  SN374  SN17081704  SN17121782  SN18011800  SN18081862  SN18061846  SN18111907",
      "more_code_info": ""
    }
  ]
}