  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Hamburg",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90490",
      "recalling_firm": "Philips Medical Systems DMC GmbH",
      "address_1": "Development And Manufacturing Ctr.",
      "address_2": "Rontgenstr. 24-26",
      "postal_code": "N/A",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-1328-2022",
      "product_description": "DigitalDiagnost C90 High Performance.  radiography and fluoroscopy system",
      "product_quantity": "140 systems in total, (Updated 1/30/2023).",
      "reason_for_recall": "Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy.   (Updated 1/30/23) Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.",
      "recall_initiation_date": "20210519",
      "center_classification_date": "20220712",
      "report_date": "20220720",
      "code_info": "Model 712034  Serial Numbers   10001073  10001104  10001108  10001109  10001119  10001126  10001135  10001140  10001144  10001145  10001153  10001156  10001157  10001163  10001170  10001173  10001178  10001179  10001183  10001191  10001192  10001199  10001200  10001201  10001203  10001208  10001210  10001216  10001220  10001231  10001232  10001234  10001235  10001236  10001240  10001241  10001248  10001251  10001261  10001262  10001264  10001270  10001271  10001273  10001275  10001280  21861349  21861369  21861402  21861418  21861456  21861460  21861471  21861475  21861490  21861496  21861497  21861498  21861501  21861502  21861525  21861531  21861542  22860038  22860118",
      "more_code_info": ""
    }
  ]
}