  "results": [
    {
      "status": "Ongoing",
      "city": "Hamburg",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84773",
      "recalling_firm": "Philips Medical Systems Gmbh, DMC",
      "address_1": "Development And Manufacturing Ctr.",
      "address_2": "Rontgenstr. 24",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution \u0013 US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MI, MN, MO, MT, NC, NJ, NY, OH, TX, UT, VA and countries of Canada, Australia, Austria, Bangladesh, Belgium, Bulgaria, Chile, China, Colombia, C¿te D'Ivoire, Cyprus, Czech Republic, Dominican Republic, Egypt, Ethiopia, France, Germany, Ghana, Hungary, Iran, Italy, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya,   Luxembourg, Malaysia, Martinique, Mexico, Nepal, Netherlands, New Zealand, Nigeria, Oman, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Viet Nam.",
      "recall_number": "Z-1328-2020",
      "product_description": "CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.",
      "product_quantity": "116",
      "reason_for_recall": "When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.",
      "recall_initiation_date": "20191219",
      "center_classification_date": "20200220",
      "termination_date": "20200618",
      "report_date": "20200226",
      "code_info": "(domestic only)  SN18061848  SN17071689  SN18051832  SN190251  SN17111759  SN17111752  SN18011793  10000020  SN17121783  SN17121785  SN18021816  SN18101890  SN19031950  SN17071688  SN18101903  SN17071683  SN17101722  SN17101729  SN17111758  SN17111763  SN17031622  SN17071692  SN17041642  SN18101886  SN18011794  SN19021934  SN19021943  SN17111770  SN18011802  SN19021942  SN17121779  SN17031618  SN17101733  SN17021613  SN17021613",
      "more_code_info": ""
    }
  ]
}