{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73561",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "N/A",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of MA, VA, TN, GA, WA, IA, NJ, TN, PA, NY, CO, CT, VT, and the countries of  Canada and Switzerland.",
      "recall_number": "Z-1327-2016",
      "product_description": "Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E    Product Usage:  The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.",
      "product_quantity": "150 units distributed in US, 1439 units distributed OUS (foreign)",
      "reason_for_recall": "Out of specification at the thread. The thread height of these screws is under-sized. The cross section of the affected area is less than that of conforming screws such that the thread height of the screws is undersized.",
      "recall_initiation_date": "20160223",
      "center_classification_date": "20160405",
      "termination_date": "20160908",
      "report_date": "20160413",
      "code_info": "Titanium Low Profile Neuro Screw, Self-Drilling, 3mm, part number # 400.833E, Lot numbers: 9958565, 9951621, 9952976, 9955377, 9962314, 9968868, 9969645",
      "more_code_info": ""
    }
  ]
}