{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Bedford",
      "address_1": "35 Crosby Dr",
      "reason_for_recall": "The audible alarm to alert the operator when cumulative irradiation time reaches five minutes is not accurate and the total cumulative irradiation time is not accurate.",
      "address_2": "",
      "product_quantity": "2780",
      "code_info": "Serial numbers per recall strategy.",
      "center_classification_date": "20130516",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) and countries of: DE, RU, AU, IT, GB, ES, FR, KR, ZA, CA, AR, LT, TW, CN, BE, AT, JP, QA, SA, SG, HK, SI, NL, and SE.",
      "state": "MA",
      "product_description": "Fluoroscan Mini C-arm InSight2, software versions 3.x, 4.0, 4.0.1, 4.0.2    The device is a Mini C-arm Fluroscopic Imaging System designed to provide physicians with general fluoroscopic visualization of a patient, including but no limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.",
      "report_date": "20130522",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hologic, Inc.",
      "recall_number": "Z-1327-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65028",
      "termination_date": "20140821",
      "more_code_info": "",
      "recall_initiation_date": "20130425",
      "postal_code": "01730-1450",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}