      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87418",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "International distribution in the country of China.",
      "recall_number": "Z-1326-2021",
      "product_description": "Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. (CHINA only)    SureSigns VM4 (863063, 863085)  SureSigns VM6 (863064, 863065, 863086)  SureSi\"gns VM8 (863066, 863068, 863087, 863088)    Model Numbers: 863063,863064,863065,863066,863068,863085,863086, 453564021961,453564021971, 53564022001,453564022041, 53564022081,453564024371, 453564024401,453564024421  453564024441,453564024461, 453564024481",
      "product_quantity": "31773 units",
      "reason_for_recall": "Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3.  Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit",
      "recall_initiation_date": "20210323",
      "center_classification_date": "20210330",
      "termination_date": "20230928",
      "report_date": "20210407",
      "code_info": "Software version: SW A.02.63",
      "more_code_info": ""
    }
  ]
}