  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mequon",
      "address_1": "6140 W Executive Dr",
      "reason_for_recall": "The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that involved patients with dense bones.  As a result of this malfunction, Titan Spine, LLC is removing all of the TCS Straight Hex lobular #6 Drivers (5210-1004) Lot# G150101 from the field.",
      "address_2": "Suite A",
      "product_quantity": "26",
      "code_info": "G150101",
      "center_classification_date": "20150326",
      "distribution_pattern": "US Distribution to the states of : AZ, CA, CT, FL, GA, ID, IL, KY, MD, MO, NV, NY, OH, OK, OR, TX, UT and WI.",
      "state": "WI",
      "product_description": "Endo TCS Driver Straight Hexalobe #6 P/N 5210-1004, Non-Sterile, Manufactured by Titan Spine, LLC.    Endo TCS driver is one of the instruments within the surgical tray for use with the Interbody Fusion Device(IBD).",
      "report_date": "20150401",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "TITAN SPINE, LLC",
      "recall_number": "Z-1324-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70576",
      "termination_date": "20150417",
      "more_code_info": "",
      "recall_initiation_date": "20150220",
      "postal_code": "53092-4499",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}