  "results": [
    {
      "country": "United States",
      "city": "Austin",
      "address_1": "7000 W William Cannon Dr",
      "reason_for_recall": "The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).",
      "address_2": "",
      "product_quantity": "15 units",
      "code_info": "Serial Numbers: CA0N0000FT, CA0N0000H0, CA0N0000H7, CA0N0000J6, CA0N0000KJ, CA0N0000KK, CA0N0000KL, CA0N0000FW, CA0N0000G5, CA0N0000G6, CA0N0000H1, CA0N0000FR, CA0N0000FZ, CA0N0000KF, CA0N0000GH",
      "center_classification_date": "20150325",
      "distribution_pattern": "US Nationwide Distribution in the states of TN, IA, AZ, NE, SD, NY, and OK.",
      "state": "TX",
      "product_description": "Coblator II System, 120V.     Product Usage:  The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgical system designed for use in otorhinolaryngology (ENT) surgery. The system consists of the following components: an electrosurgical radiofrequency controller, a reusable, non-sterile Power Cord; a reusable, non-sterile Foot Control; a reusable, non-sterile Flow control Valve Unit; a reusable, non-sterile flow control cable; a single use, disposable, sterile Wand. The Coblator II is indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery.",
      "report_date": "20150401",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "ArthroCare Medical Corporation",
      "recall_number": "Z-1323-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70594",
      "termination_date": "20160226",
      "more_code_info": "",
      "recall_initiation_date": "20150220",
      "postal_code": "78735-8509",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}