  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Atlanta",
      "address_1": "400 Perimeter Center Ter Ne",
      "reason_for_recall": "The problem is that the \"static tolerances\" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values.",
      "address_2": "Ste 50",
      "product_quantity": "498",
      "code_info": "152718, 152003, 152007, 151963, 152081, 151693, 152131, 152318, 105792, 105570, 152778, 151052, 152363, 152472, 152195, 153022, 152689, 152640, 152642, 152663, 152680, 152782, 152796, 152829, 152862, 152962, 153030, 153125, 153244, 153270, 153211, 153296, 152834, 152596, 151354, 151353, 152664, 151200, 151823, 151822, 151610, 152026, 152027, 152692, 151276, 152856, 151106, 152877, 152986, 152810, 105935, 153105, 153069, 151023, 152914, 151975, 151240, 151239, 153135, 105810, 153145, 153129, 151182, 151650, 152886, 151850, 152831, 153132, 151343, 153063, 105985, 152832, 152845, 152837, 152863, 152859, 152776, 152857, 152866, 152876, 152820, 152870, 152884, 152865, 152872, 152909, 152910, 152911, 152918, 152934, 152961, 152966, 153009, 153041, 153026, 153078, 153079, 153083, 153070, 153066, 153027, 153154, 153157, 153159, 153165, 153131, 153352, 152675, 152737, 152299, 152403, 152676, 151670, 152252, 151153, 151482, 151296, 152068, 152286, 152097, 151554, 151768, 151853, 151854, 151786, 151495, 151787, 153099, 152935, 151865, 151466, 105920, 152521, 151536, 151263, 105988, 151028, 151058, 152719, 151174, 151597, 152645, 152650, 152647, 105755, 152743, 151857, 151921, 105896, 151986, 152855, 152092, 151006, 151277, 105314, 105919, 152746, 152770, 105901, 151635, 152575, 152627, 151701, 151922, 151852, 152422, 152309, 152498, 152628, 152706, 152616, 105850, 152014, 151688, 151638, 152721, 153234, 152816, 105554, 105987, 151452, 151166, 152913, 151740, 152042, 152904, 152254, 152753, 152554, 152591, 152644, 152607, 152648, 152687, 152714, 152899, 152739, 152672, 152900, 152952, 152977, 153080, 153169, 153183, 152908, 153349, 153325, 152702, 152869, 152724, 151529, 152840, 152841, 151442, 151445, 151612, 151613, 151756, 105677, 151755, 151189, 151034, 151652, 152757, 105893, 105892, 152945, 151060, 105711, 105997. 151324, 151316, 151318, 151385, 152873, 151660, 152734, 152646, 152682, 151087, 151088, 105833, 152613, 152958, 105946, 151117, 151308, 151108, 151280, 151912, 152106, 152157, 152294, 152637, 152624, 152667, 152727, 152726, 152750, 152893, 152953, 151132, 151150, 151879, 151583, 152595, 152733, 153087, 152975, 153310, 153229, 152738, 152735, 152785, 152797, 152947, 153327, 152907, 153291, 153290, 152677, 152674, 151112, 105937, 151349, 105838, 105758, 105927, 151045, 151314, 151553, 151140, 151101, 105995, 105915, 151405, 152638, 152325, 152457, 152512, 151196, 151368, 152118, 152119, 105968, 151040, 105912, 151247, 105943, 151005, 105948, 151341, 151898, 151703, 151982, 152161, 151718, 153059, 151297, 151278, 151020, 105928, 152603, 153084, 152942, 151829, 151284, 152298, 151449, 151144, 151602, 152708, 152713, 152605, 152606, 152614, 152611, 152629, 152679, 152799, 152880, 152928, 152920, 152336, 153343, 153348, 152101, 151669, 151237, 151121, 153205, 151897, 105415, 152793, 105640, 105703, 105869, 151175, 151066, 151233, 152060, 151503, 152190, 153333, 152795, 151089, 152660, 152538, 152827, 152848, 152442, 152846, 152723, 152921, 151119, 152769, 152968, 152143, 152720, 151497, 151197, 151420, 105721, 151079, 105787, 105685, 151511, 151022, 105743, 105742, 151249, 151042, 105786, 151516, 151109, 151331, 151524, 152592, 151255, 105813, 151556, 151526, 105811, 151032, 151581, 151243, 151492, 152070, 151096, 151004, 151136, 105797, 105802, 105957, 105789, 151292, 151425, 151291, 152148, 151332, 151333, 151486, 151415, 151573, 151528, 151487, 151527, 151572, 151414, 105874, 151225, 152653, 105814, 151092, 151047, 151706, 151691, 151692, 151934, 151945, 151946, 153190, 151127, 105412, 152030, 152168, 152396, 152654, 151749, 152891, 152878, 152858, 105894, 151033, 151258, 151168, 151159, 151228, 152760, 151791, 152522, 151391, 151216, 151539, 151611, 151696, 151867, 152965, 151325, 152220, 151785, 152217, 151388, 151394, 151892, 152043, 151793, 151684, 152223, 152303, 152358, 151683, 152652, 152659, 152703, 151672, 152767, 152777, 152783, 152784, 152787, 152798, 152881, 151710, 153045, 152929, 153137.",
      "center_classification_date": "20140402",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CT, FL, GA, KY, LA, MD, MO, NM, NY, NC, OH, OK, PA, TN, TX, and WA and the countries of Argentina, Australia, Austria, Belarus, Belgium, Brazil, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Poland, Romania, Russia, Saudi Arabia, Senegal, Singapore, Slovakia, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.",
      "state": "GA",
      "product_description": "Integrity 1.1.    Integrity is the interface and control software for the Elekta range of medical digital linear accelerators and is intended to assist a licensed practitioner in the delivery of radiation to defined target volumes.",
      "report_date": "20140409",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Elekta, Inc.",
      "recall_number": "Z-1322-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67574",
      "termination_date": "20170726",
      "recall_initiation_date": "20130328",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}