{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.",
      "address_2": "",
      "product_quantity": "12543",
      "code_info": "Serial Numbers: All  Including UDIs: (01)00884838028890(21) and (01)00884838087075(21) followed by serial number.",
      "center_classification_date": "20190513",
      "distribution_pattern": "Worldwide distribution.  US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea Republic, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Martinique, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Philippines, Portugal, Qatar, Reunion, Romania, Rwanda, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Kingdom,  United Arab Emirates,Venezuela, Vietnam, and Zimbabwe.",
      "state": "MA",
      "product_description": "SureSigns VS3 NBP, SpO2, Product Number 863071.  For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).",
      "report_date": "20190522",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips North  America, LLC",
      "recall_number": "Z-1319-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "82684",
      "more_code_info": "",
      "recall_initiation_date": "20190419",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}