  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Batesville",
      "address_1": "1069 State Road 46 E",
      "reason_for_recall": "A component used within the braking system may break resulting in one or more brake pedals on the bed staying in brake or not being engaged which could result in unwanted bed movement and possible patient or user injury.",
      "address_2": "",
      "product_quantity": "2117",
      "code_info": "S159AW9192 through S305AW2321  Beds manufactured from June 8, 2017 to November 1, 2017",
      "center_classification_date": "20180409",
      "distribution_pattern": "Worldwide distribution.  US nationwide, Canada, Chile, Colombia, Costa Rica, France, Hong Kong, Israel, Jamaica, Japan, Jordan, Malaysia, Mexico, Oman, Panama, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, UAE, and UK.",
      "state": "IN",
      "product_description": "Progressa Bed -  AC-powered adjustable hospital bed",
      "report_date": "20180418",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hill-Rom, Inc.",
      "recall_number": "Z-1319-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79516",
      "termination_date": "20200626",
      "more_code_info": "",
      "recall_initiation_date": "20180112",
      "postal_code": "47006-7520",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}