  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saint Paul",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67716",
      "recalling_firm": "3M Company - Health Care Business",
      "address_1": "3M Center 2510 Conway Ave , B# 275-5-W-6",
      "address_2": "N/A",
      "postal_code": "55144",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide)  Distribution including the states of AL, AZ, CA,CO, FL, GA, HI, IL, IN, IA, KS, KY, MD, MA, MI, MN, MS, NE, NJ, NY, NC, ND, OH, OK,OA, SC, TX, WA, WV and WI. and the countries of COLOMBIA, VENEZUELA, UAE, THAILAND, GERMANY, CHINA, KOREA, CANADA, JAPAN, RUSSIA, INDIA, POLAND, PERU, AUSTRIA, ISRAEL, INDONESIA, SPAIN, CHILE, MALAYSIA, SWITZERLAND, EL SALVADOR, ECUADOR, LEBANON, MEXICO, ARGENTINA, BULGARIA, TAIWAN, TURKEY,ROMANIA, GUATEMALA and NEW ZEALAND.",
      "recall_number": "Z-1318-2014",
      "product_description": "3M\" Steri-Vac\" Gas Sterilizer/Aerators, Model #'s 4XL,5XL, 8XL    The 3M\" Steri-Vac\" Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- and/or moisture-sensitive devices. This gas sterilizer/aerator is intended for INDOOR USE ONLY.",
      "product_quantity": "73 US, 346 OUS",
      "reason_for_recall": "Recently, 3M received a single report of a potential exposure to Ethylene Oxide (EO) at a hospital installation involving 3M\" Steri-Vac\" Gas Sterilizers. The cause was traced to a rare situation involving a failure of the EO cartridge puncture assembly. This assembly was only in production during the period November 10, 2006 through December 14, 2007 and in service during the period February 7, 20",
      "recall_initiation_date": "20140317",
      "center_classification_date": "20140402",
      "termination_date": "20150122",
      "report_date": "20140409",
      "code_info": "Model SKU            Starting S/N  Ending S/N  5XL 70-2007-2303-2 721115     721309  5XL 70-2007-2304-0 820174     820201  5XL 70-2007-2575-5 780066     780102  8XL 70-2006-8493-7 350136     350268  8XL 70-2006-8498-5 450016     450027    4XL 523122, 522205, 522502, 522287, 522655."
    }
  ]
}