  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Memphis",
      "address_1": "1450 E Brooks Rd",
      "reason_for_recall": "One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.",
      "address_2": "",
      "product_quantity": "59 units",
      "code_info": "Lot Number 17GM04522",
      "center_classification_date": "20180406",
      "distribution_pattern": "US, Puerto Rico, Malaysia",
      "state": "TN",
      "product_description": "D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R",
      "report_date": "20180418",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Smith & Nephew, Inc.",
      "recall_number": "Z-1317-2018",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "79515",
      "termination_date": "20181031",
      "more_code_info": "",
      "recall_initiation_date": "20180307",
      "postal_code": "38116-1804",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}