  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Feldkirchen",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70657",
      "recalling_firm": "Brainlab AG",
      "address_1": "Kapellenstrasse 12",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - USA and to the countries of : Australia, Austria, Canada Denmark, France, Germany, Hong Kong, Japan, Russia, Singapore, South Korea and United Kingdom.",
      "recall_number": "Z-1316-2015",
      "product_description": "ExacTrac versions 6.x patient positioning systems are used to position patients during radiosurgery or radiotherapy procedures..",
      "product_quantity": "24 systems (US); 36 systems (Foreign)",
      "reason_for_recall": "The ExacTrac 6.x Patient Positioning System may incorrectly position the patient when using the ExacTrac Cone Beam CT (CBCT) with a TrueBeam-specific optional subvolume-CBCT.",
      "recall_initiation_date": "20150216",
      "center_classification_date": "20150325",
      "termination_date": "20170228",
      "report_date": "20150401",
      "code_info": "ExacTrac versions 6.x  including versions 6.0.0, 6.0.1, 6.0.2, 6.0.3, 6.0.4, 6.0.5; and v. 6.1.0.     Model/Catalog Numbers:     49971, CBCT IMPORT & ALIGNMENT SOFTWARE;   20836, EXACTRAC 6.0 VARIAN CBCT LICENSE;     The software may be contained in configured items:     48320, ET UPGRADE SOFTWARE 6.0 TO 6.1;   48320-99, ET UPGRADE SOFTWARE 6.0 TO 6.1 SLA;   49973B, ET DATA PREP / REVIEW SYSTEM;   49973A, ET DATA PREP / REVIEW SYSTEM;   49973, ET DATA PREP / REVIEW SYSTEM;   49998B, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;   49998A, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;   49998, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT;   48304, ET 6.1 DATA PREP/REVIEW SYSTEM UPG. KIT;   48302, ET DATA PREP / REVIEW SYSTEM",
      "more_code_info": ""
    }
  ]
}