  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Sparks",
      "address_1": "7 Loveton Cir",
      "reason_for_recall": "Background calibration has the potential to cause false positive or false negative test results. An incorrect test result with the BD Veritor System could lead to possible misdiagnosis and inappropriate treatment",
      "address_2": "",
      "product_quantity": "32 units",
      "code_info": "Serial Numbers:  Serial Number  1807138J1E050  1710067J160C0  1703027J07FD0  1810268J2D370  1710217J18470  1703027J084C0  1703307J0C630  1710067J16190  1709187J12360  1710267J19D40  1710067J16230  1710067J161D0  1710067J161F0  1710067J16260  1807138J1E0B0  1609076J04270  1802138J07B40  1803298J11070  1802078J05D80  1704067J0DA00  1612166J070C0  1710257J18F50  1802028J044C0  1703167J0B0B0  1612026J07010  1710067J16110  1802058J05210  1811128J2FF50  1710067J16200  1710067J16270  1710237J185C0  1702147J04360",
      "center_classification_date": "20190510",
      "distribution_pattern": "Nationwide  Foreign:  Canada, Chile, Hong Kong, Japan, Taiwan",
      "state": "MD",
      "product_description": "BD Veritor Plus Analyzer- In-Vitro Diagnostic  intended to use with the BD Veritor System Test Devices.  Catalog Number: 256066",
      "report_date": "20190522",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Becton Dickinson & Co.",
      "recall_number": "Z-1313-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "82616",
      "termination_date": "20200504",
      "more_code_info": "",
      "recall_initiation_date": "20190403",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}