  "results": [
    {
      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79462",
      "recalling_firm": "Roche Diagnostics Corporation",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US Distribution in states of: CA, FL and IN.",
      "recall_number": "Z-1311-2018",
      "product_description": "Elecsys BRAHMS PCT    Immunoassay for the in vitro quantitative determination of PCT (procalcitonin) in human serum and plasma (K2 and K3 EDTA, Li-Heparin).",
      "product_quantity": "47",
      "reason_for_recall": "Product exhibits a decreased recovery of patient samples on the cobas e 801 module compared to the MODULAR ANALYTICS E 170, cobas e 411, 601, and 602 analyzers.",
      "recall_initiation_date": "20171222",
      "center_classification_date": "20180404",
      "termination_date": "20190405",
      "report_date": "20180411",
      "code_info": "Elecsys BRAHMS PCT: 07301715190 Lot number 204084"
    }
  ]
}