  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Grand Rapids",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90400",
      "recalling_firm": "Skytron, Div. The KMW Group, Inc",
      "address_1": "5085 Corporate Exchange Blvd SE",
      "address_2": "",
      "postal_code": "49512-5515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada and Japan.",
      "recall_number": "Z-1309-2022",
      "product_description": "Product Name: GCX Channel Mounting Accessory  Model/Catalog Number: H8-300-40-A",
      "product_quantity": "731",
      "reason_for_recall": "Under certain circumstances, the mounting screws may loosen or pull out from the attachment point on the Ergon 3 boom when the GCX Channel Mounting Accessory was used with the VST clamp as a monitor mount on the boom.",
      "recall_initiation_date": "20220602",
      "center_classification_date": "20220630",
      "report_date": "20220706",
      "code_info": "Lot Code: All units of model H8-300-40-A are affected. Code Info: GCX mounting bracket does not have UDI. Ergon 3 booms after September 2016 have UDI. DI of units manufactured after September 2016 is 10841736100889. Units manufactured prior to September 2016 do not have UDI.",
      "more_code_info": ""
    }
  ]
}