  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Oakwood Village",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73550",
      "recalling_firm": "Viewray Incorporated",
      "address_1": "2 Thermo Fisher Way",
      "address_2": "N/A",
      "postal_code": "44146-6536",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US including CA, FL, MO & WI., and International distribution to the Republic of Korea and The Netherlands.",
      "recall_number": "Z-1309-2016",
      "product_description": "MRIdian ViewRay Radiation Therapy System, ViewRay Treatment Planning and Delivery System (also known as the MRIdian¿ System) is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated",
      "product_quantity": "5 units",
      "reason_for_recall": "When editing the isocenter or the couch position of the plan while in the treatment workflow (in the Points screen) and re-optimizing, the software will not prompt the user to shift the couch to the new isocenter.  As  a result there is the potential to deliver dose to the initial isocenter rather than the new location.",
      "recall_initiation_date": "20160307",
      "center_classification_date": "20160331",
      "termination_date": "20170222",
      "report_date": "20160406",
      "code_info": "Model number: 10000  Catalog/Part Number: 10000  Serial Numbers: 100, 101, 102, 104 & 105"
    }
  ]
}