{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Pittsburgh",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79433",
      "recalling_firm": "Dynavox Systems Llc",
      "address_1": "2100 Wharton St Ste Ste 400",
      "address_2": "N/A",
      "postal_code": "15203-1691",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationally",
      "recall_number": "Z-1305-2018",
      "product_description": "Dynavox T15 Speech Generating Device ( including Batteries)",
      "product_quantity": "307",
      "reason_for_recall": "The battery is integral to the device.  If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire.  The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells.  This battery may cause the device case to separate a small amount.",
      "recall_initiation_date": "20171214",
      "center_classification_date": "20180403",
      "termination_date": "20191211",
      "report_date": "20180411",
      "code_info": "TL1405   TL1407   Part Numbers: T15: 750423  Date Codes are part of the device s unique Serial Number and represent the year and month the device was manufactured."
    }
  ]
}