{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94021",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the country of Taiwan.",
      "recall_number": "Z-1304-2024",
      "product_description": "DCA Systems HbA1c Reagent Kit",
      "product_quantity": "74,211 Kits (Expanded 2/28/24: additional 72,892 kits)",
      "reason_for_recall": "The reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.",
      "recall_initiation_date": "20240207",
      "center_classification_date": "20240319",
      "report_date": "20240327",
      "code_info": "Siemens Material Number: 10311134; UDI/DI: 00630414532806; Lot#: 9640093, 0645103, 0648103, 0649103, 0654103 . Expanded Lots (02/28/2024): 0637093, 0641093, 0642093, 0647103, 0653103, 0655103;",
      "more_code_info": ""
    }
  ]
}