  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Sunnyvale",
      "address_1": "1266 Kifer Rd Bldg 100",
      "reason_for_recall": "Correction due to a fault that causes distal SUJ (DSUJ) failures that indicates a mismatch between the positions reported by the redundant position sensors on one of the patient cart arm joints.",
      "address_2": "",
      "product_quantity": "167 total da Vinci Xi  affected systems",
      "code_info": "Material number: 3680662-20:  Serial numbers: 1311140273, 1311140370, 1311140385, 1311140148, 1311140099, 1311140152,  1311140081, 1311140473, 1311140561, 1311140409.   Material number: 380662-21:  Serial numbers: 1311140225, 1311140437, 1311140476, 1311140336,   1311140472, 1311140438, 1311140353, 1311140195, 1311140545, 1311140098  1311140189, 1311140465, 1311140394, 1311140366, 1311140008, 1311140115  1311140014, 1311140371, 1311140238, 1311140306, 1311140332, 1311140404  1311140054, 1311140188, 1311140324, 1311140453, 1311140408, 1311140237  1311140220, 1311140372, 1311140325, 1311140088, 1311140535, 1311140355.",
      "center_classification_date": "20150320",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Australia, Belgium, China, France, Germany, India, Israel, Italy, Norway, South Korea, Spain, Sweden, Switzerland, Turkey, and United Kingdom.",
      "state": "CA",
      "product_description": "da Vinci Xi Surgical System, IS4000;  Assy, Distal SUJ, Inner, IS4000  Material number: 380662-20 and 380662-21.     The Intuitive Surgical Endoscope Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endocopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders,  endscopic retractors, electrocautery and accessories for edoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximations, ligation, electrocautery, suturing and delivery and placement of microware and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic labaroscopic surgical  procedures, general thorascopic surgical procedures and thorascopically-assisted cardiotomy procedures.",
      "report_date": "20150401",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Intuitive Surgical, Inc.",
      "recall_number": "Z-1302-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70705",
      "termination_date": "20150908",
      "more_code_info": "",
      "recall_initiation_date": "20150303",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}