  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Birmingham",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79499",
      "recalling_firm": "The Binding Site Group, Ltd.",
      "address_1": "8 Calthorpe Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution was made to CA.  There was no foreign/military/government distribution.",
      "recall_number": "Z-1301-2018",
      "product_description": "Binding Site Optilite Beta-2-Microglobulin Urine Kit, Ref LK043.L.OPT.A, IVD.  The firm name on the label is The Binding Site Group Ltd., Birmingham, U.K.    The Optilite Beta-2-Microglobulin (¿2M) Urine kit is intended for the quantitative in-vitro measurement of beta-2-microglobulin in urine using the Binding Site Optilite analyser to aid in the diagnosis of active rheumatoid arthritis and kidney disease. This test should be used in conjunction with other laboratory and clinical findings.",
      "product_quantity": "20 kits distributed to the U.S.",
      "reason_for_recall": "The product is not meeting the antigen excess protection claim of up to 760mg/L, as stated in the Instructions for Use.",
      "recall_initiation_date": "20180309",
      "center_classification_date": "20180403",
      "termination_date": "20200622",
      "report_date": "20180411",
      "code_info": "Lot 407460, UDI 05051700017688"
    }
  ]
}