  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82621",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, and WI.",
      "recall_number": "Z-1298-2019",
      "product_description": "Cios Alpha, Model Number 10308191  Mobile X-ray system - Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.",
      "product_quantity": "257",
      "reason_for_recall": "Continuous operation at very high tube output may result in increased wear of the radiation emitting components. This may occur when tube voltages of 125 kV are used for a long period, e.g. 30 minutes or more of continuous use. As a result, risk of premature failure of the Monoblock and loss of imaging X-rays.",
      "recall_initiation_date": "20190318",
      "center_classification_date": "20190508",
      "termination_date": "20200416",
      "report_date": "20190515",
      "code_info": "Serial Numbers:  11294  11202  12006  12009  13098  13016  12823  13011  11163  13009  13059  12662  11032  11311  13007  11258  11379  11380  12530  12334  11376  12426  11336  11132  11295  13004  13121  11044  12859  11005  11333  11386  10076  12677  10115  11291  11056  11287  12319  12529  12617  12623  12657  12876  12886  13048  13050  13105  13109  13110  12848  12574  12597  12647  13002  12875  12810  10078  12306  12317  12874  11119  12576  12678  12513  12592  12490  12885  11200  11281  11283  12425  12023  13070  12396  12920  11285  11103  11105  11257  11074  12379  11013  10021  11330  11066  13012  10057  11143  11077  11319  11240  11349  13082  11098  12815  12339  11153  11372  10037  11328  13008  12308  10063  11106  12373  12584  11067  10082  12026  12027  11168  12492  10019  11317  10035  11198  12040  12307  12440  11165  12611  11094  11226  11370  12970  12377  12861  10089  11082  11080  11009  11018  10058  12321  12718  12004  12999  11378  11167  11248  12817  11306  11229  12028  11144  12313  12867  12868  12869  12870  12881  12887  12889  12894  11038  12320  12514  11352  12612  12811  12838  11243  11314  12626  11374  12899  12902  12884  11162  11292  12394  12498  12549  11120  11134  13062  13064  12494  12925  12526  11219  11220  11346  12993  12956  12562  12607  12025  12805  12046  11288  12816  11356  12620  11124  12557  12806  11251  12322  12323  12533  12504  13073  12958  11015  11016  12486  12515  12500  12835  11327  11246  12878  10096  12391  12622  12652  12891  11290  12416  12472  12496  12552  12946  10116  12578  12642  12873  12599  12813  11280  11035  12003  12732  12570  12855  13045  13049  12030  12655  12818  12819  11359  12314  12618  12809  13033  11338  12947  12948  11064  12842  12457  12460  12972  13107"
    }
  ]
}