      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67502",
      "recalling_firm": "Draeger Medical Systems, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-1298-2014",
      "product_description": "Draeger PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500),  Evita V500 and /or Babylog VN500 Ventilators, used for ventilation during short term transport of patients within a hospital.",
      "product_quantity": "1544",
      "reason_for_recall": "The battery capacity of the optional PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500) did not last as long as expected.",
      "recall_initiation_date": "20140207",
      "center_classification_date": "20140328",
      "termination_date": "20150817",
      "report_date": "20140409",
      "code_info": "all lots"
    }
  ]
}