  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Flower Mound",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87331",
      "recalling_firm": "Stryker Communications",
      "address_1": "571 Silveron Blvd",
      "address_2": "N/A",
      "postal_code": "75028",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV, and WY. The country of Canada.",
      "recall_number": "Z-1297-2021",
      "product_description": "S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502  Components that are affected within the S-Series Finished Good Model Number  o\tP38198 - ASM, MFR CONTROLLER, FIXED STANDARD  o\tP38199 - ASM, MFR CONTROLLER, LIFT STANDARD  o\tP40553 - ASM, MFR HANDLE REPAIR, LIFTING, S-SERIES, STANDARD  o\tP40554 - ASM, MFR HANDLE REPAIR, FIXED, S-SERIES, STANDARD  o\tP46269 - ASSY, MFR CONTROLLER UPGRADE, STANDARD",
      "product_quantity": "1,393 units",
      "reason_for_recall": "The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.",
      "recall_initiation_date": "20210129",
      "center_classification_date": "20210324",
      "report_date": "20210331",
      "code_info": "Serial Numbers: 2258046   2863490",
      "more_code_info": ""
    }
  ]
}