  "results": [
    {
      "status": "Ongoing",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87386",
      "recalling_firm": "Getinge Group Logistics America, LLC",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV. Foreign: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Latvia, Lebanon, Liechtenstein, Lithuania, Luxembourg, Lybia, Macedonia, Malaysia, Mauritius, Mexico, Morocco, Myanmar, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, United Kingdom, Uzbekistan, Venezuela, Vietnam and Yemen.",
      "recall_number": "Z-1296-2021",
      "product_description": "Operating Table System",
      "product_quantity": "4288 units",
      "reason_for_recall": "There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.",
      "recall_initiation_date": "20210211",
      "center_classification_date": "20210323",
      "report_date": "20210331",
      "code_info": "Part Number: 118001A0, 118001B0, 118001B1,  118001B2, 118001B3, 118001B4, 118001C0, 118001D0. UDI code: 04046768031500, 04046768031531, 04046768040700, 04046768088351, 04046768088061, 04046768124752, 04046768031555, 04046768031579. Serial Number: 1   3402, 1   1028, 1   1264, 1   401, 3   74, 1   81, 2   1159, 2 - 762.",
      "more_code_info": ""
    }
  ]
}