  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bedford",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82217",
      "recalling_firm": "Medica Corporation",
      "address_1": "5 Oak Park Dr",
      "address_2": "N/A",
      "postal_code": "01730-1430",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide distribution - US Nationwide and countries of   IRAQ, FRANCE, INDIA, GEORGIA, GEORGIA, FRANCE  GUATEMALA, KENYA, IRAN, ESTONIA.",
      "recall_number": "Z-1294-2019",
      "product_description": "Medica Wash1-Wash solution used prior to Lipase assay on the EasyRA clinical chemistry analyzer   REF 10680 - Product Usage: For use in washing the probe of the Medica EasyRA Chemistry Analyzer. The wash solution in the wedge is used to prevent interference between certain tests.",
      "product_quantity": "87 kits",
      "reason_for_recall": "Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.",
      "recall_initiation_date": "20190206",
      "center_classification_date": "20190507",
      "termination_date": "20201002",
      "report_date": "20190515",
      "code_info": "Lot #s 18066, 18133, 18235, 18317"
    }
  ]
}