  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Orange",
      "address_1": "1717 W Collins Ave",
      "reason_for_recall": "Cases of Cure Sleeve labeled with part number 4513, which fits 13mm tips, have inner boxes  labeled with part number 4500 which fit 7mm and 8mm tips.  There is a low risk associated with the use of the product as the sleeve may tear causing a potential for cross contamination.",
      "address_2": "",
      "product_quantity": "172",
      "code_info": "Part Number: 4513C or 4513-1.    Lot Number: 2014030110",
      "center_classification_date": "20150318",
      "distribution_pattern": "Worldwide distribution.  US Nationwide, Canada, Australia, Netherlands, Switzerland, Japan,",
      "state": "CA",
      "product_description": "Outer cardboard box labeling:  Pinnacle Cure Sleeve, Curing Light Tip Sleeves.    Kerr Total Care.  Part Number: 4513C or 4513-1  Single Patient Use  Fits 13mm Tips. Pack of 400.      Metrex Research, Manufacturer.    Inner individual packaging of each sleeve:  Cure Sleeve by Pinnacle  Model 4500",
      "report_date": "20150325",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Sybron Dental Specialties",
      "recall_number": "Z-1291-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70551",
      "termination_date": "20150521",
      "more_code_info": "",
      "recall_initiation_date": "20150211",
      "postal_code": "92867-5422",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}