  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Reykjavik",
      "state": "",
      "country": "Iceland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70648",
      "recalling_firm": "Ossur H / F",
      "address_1": "Grjothals 5",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide within the U.S. and internationally to the following foreign countries; Belgium, United Kingdom, Germany, Czech Republic, France, Israel, Poland, Spain, Australia, Japan, Mexico, Bulgaria, Canada, South Korea, Sweden, Finland,",
      "recall_number": "Z-1288-2015",
      "product_description": "LP Rotate Foot System  prosthetic device for fitting lower extremity amputations.   -Low profile foot module with adjustable shock absorber and rotation  with progressive stiffening  -Build height: with male pyramid adapter 145 - 153mm (5 11/16  - 6\u001d )  with female pyramid adapter 153  165mm (6 - 6 1/2\u001d)  -Male pyramid adapter  -Female pyramid adapter  -Maintenance free, no lubrication required  -Maximum weight: 147kg (325lbs)  -Available in sizes 22-30",
      "product_quantity": "2,194 (1,464 in U.S. and 730 internationally)",
      "reason_for_recall": "The threads of the outer tube of the shock module of LP Rotate are too short, resulting in insecure assembly. Use of the product may cause the patient to fall with continued use, as the shock module can become loose from the foot blade.",
      "recall_initiation_date": "20150304",
      "center_classification_date": "20150317",
      "termination_date": "20151028",
      "report_date": "20150325",
      "code_info": "LP Rotate, sizes 25-27 cat.1-6. Product numbers LRP062, LRF062, LRPU62, LRFU62. All from manufacturing lot number HF141022",
      "more_code_info": ""
    }
  ]
}