  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Saint Paul",
      "address_1": "1265 Grey Fox Rd",
      "reason_for_recall": "Smiths Medical has become aware of an issue with specific lots of 50mL CADD\" Medication Cassette Reservoirs (\u001cCassette\u001d). Some Cassettes may leak at the sealing area of the pump tube and medication bag. Smiths Medical has received no reports of  serious injury or death related to this issue.",
      "address_2": "",
      "product_quantity": "48,208 (17,040 US, 31,168 OUS).",
      "code_info": "21-7001-24: Lots 14X297 and 14X323  21-7301-24 (Flow Stop): Lot 14X324",
      "center_classification_date": "20150317",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) Distribution to the states of : AL, AZ, AR, CA, CO, CT, FL, GA, HI, IL IN, IA, MD, MA, MN, MS, MO, NE, NV, NH ,NJ, NM, NY, NC, OH, OR, PA, SD, TN, TX, UT, VA and WA., and to the countries of : Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Czech Republic, Germany, Denmark, Spain. Finland, France, United Kingdom (GB) Hungary, Indonesia, Ireland, Italy, Netherlands, Portugal, Sweden, Singapore and South Africa.",
      "state": "MN",
      "product_description": "CADD\" Medication Cassette Reservoir, 50mL,  REF 21-7001-24 & 21-7301-24, Sterile EO, Rx  only.",
      "report_date": "20150325",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Smiths Medical ASD, Inc.",
      "recall_number": "Z-1286-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70402",
      "termination_date": "20160318",
      "more_code_info": "",
      "recall_initiation_date": "20150202",
      "postal_code": "55112-6929",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}