  "results": [
    {
      "status": "Terminated",
      "city": "Pleasanton",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87445",
      "recalling_firm": "CAIRE DIAGNOSTICS INC",
      "address_1": "7020 Koll Center Pkwy",
      "address_2": "N/A",
      "postal_code": "94566-3103",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution in the states of CA, CO, FL, TX, and WI. In the countries of Belgium, Canada, Croatia/Hrvatska, France, Germany, Greece, Hong Kong, India, Philippines, Spain, Spain and Switzerland.",
      "recall_number": "Z-1284-2021",
      "product_description": "CAIRE DIAGNOSTICS FenomPro REF 900-0001 SN ***** - Product Usage: a portable, non-invasive device to measure fractional exhaled nitric oxide (FeNO) in human breath. Fenom Pro should not be used in critical care, emergency care or in anesthesiology.",
      "product_quantity": "381 units (171 units U.S. and 210 units O.U.S.)",
      "reason_for_recall": "Erroneously high FeNO levels which may contribute to premature discontinuation of effective asthma treatment and persistent underlying disease.  Two issues affecting the FeNO readings:  1) Drift in the calibration gases and 2) Software error.  Each of these issues can cause FeNO scores/results to be erroneously high.",
      "recall_initiation_date": "20210226",
      "center_classification_date": "20210319",
      "termination_date": "20230324",
      "report_date": "20210331",
      "code_info": "All units sold before February 23, 2021 are affected (serial numbers 100107 to 100651).",
      "more_code_info": ""
    }
  ]
}