{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73672",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY including Puerto Rico.",
      "recall_number": "Z-1284-2016",
      "product_description": "AXIOM Artis zee/zeego, system, x-ray, angiographic    Product Usage:  AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.",
      "product_quantity": "986 systems in total",
      "reason_for_recall": "Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet.  Loss of coolant can result in a cooling system failure and subsequent system failure.  This leak can occur sporadically in affected systems.",
      "recall_initiation_date": "20160218",
      "center_classification_date": "20160331",
      "termination_date": "20170808",
      "report_date": "20160406",
      "code_info": "Serial Numbers:   Artis Q Ceiling      109033, 109135, 109141, 109123, 109080, 109074, 109119, 109016, 109118, 109155, 109067, 109066, 109068, 109134, 109133, 109084, 109011, 109096, 109159, 109130, 109143, 109102, 109015, 109144, 109009, 109057, 109108, 109103, 109038, 109040, 109034, 109129, 109137, 109029, 109136, 109064, 109132, 109058, 109145, 109050, 109076, 109007, 109072, 109075, 109071, 109091, 109090, 109041, 109039, 109060, 109146, 109088, 109092, 109032, 109021, 109048, 109020, 109036, 109043, 109161, 109101, 109105, 109104, 109107, 109047, 109031, 109028, 109027, 109120, 109010, 109093, 109157, 109156, 109069, 109158, 109044"
    }
  ]
}