  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Plainsboro",
      "address_1": "105 Morgan Ln",
      "reason_for_recall": "Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.",
      "address_2": "",
      "product_quantity": "21,179 boxes",
      "code_info": "1125523, 1125517  1125522, 1125539  1125524, 1125525  1125526, 1125632  1125630, 1125678  1125679  1125677, 1125683  1125628, 1125813  1125814, 1125815  1125629, 1125520  1125521, 1125979  1126031, 1125980  1125981, 1125982  1126030, 1130243  1130071, 1130503  1130379, 1130381  1130384, 1130507  1130374, 1130380  1130382, 1130421  1130505, 1130424  1130654, 1130422  1130423, 1130428  1130639, 1130413  1130414, 1130383  1130425, 1130504  1130506, 1130426  1130427, 1130970  1130827, 1111606  1104879, 1110814  1111610, 1104876  1110815, 1111607  1110268, 1111360  1112339, 1104405  1104877, 1105455  1112008, 1112110  1112111, 1110845  1111940, 1112149  1104880, 1112148  1110244, 1110265  1110846, 1104878  1111609, 1111941  1104881, 1110812,  1110267, 1111359,  1112340, 1105307,  1111611, 1111939,  1112341, 1110266  1112146, 1112147",
      "center_classification_date": "20130510",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Britain, Italy, Japan, Peru, Singapore, Thailand, and South Africa.",
      "state": "NJ",
      "product_description": "DuraGen Plus Dural Regeneration Matrix    DuraGen Plus Adhesion Barrier Matrix is an absorbable implant for repair and restoration of dural defects and as an adhesion barrier for the reduction of peridural fibrosis. DuraGen Plus matrix is an easy to handle, soft, white pliable, non-friable, porous, collagen matrix. DuraGen Plus matrix is supplied sterile, nonpyrogenic, for single use in double peel packages in a variety of sizes.",
      "report_date": "20130522",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Integra LifeSciences Corporation",
      "recall_number": "Z-1284-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64778",
      "termination_date": "20150603",
      "more_code_info": "",
      "recall_initiation_date": "20130328",
      "postal_code": "08536-3339",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}