  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84769",
      "recalling_firm": "Covidien LLC",
      "address_1": "60 Middletown Ave",
      "address_2": "",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1282-2020",
      "product_description": "Polysorb Braided Absorbable Suture 1 36 VIOLET GS-24, Item Code CL915 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in cardiovascular or neural tissue.",
      "product_quantity": "139,938 units total",
      "reason_for_recall": "There is a potential for packaging integrity issues impacting the humidity or sterility barrier of specific production lots.",
      "recall_initiation_date": "20191218",
      "center_classification_date": "20200219",
      "termination_date": "20220816",
      "report_date": "20200226",
      "code_info": "1) Lot - A9H0927Y  Expiration Date - 2024-07-31;     2) Lot - A9H1187Y  Expiration Date - 2024-07-31;     3) Lot - A9J0383Y  Expiration Date - 2024-08-31;     4) Lot - A9J0428Y  Expiration Date - 2024-08-31",
      "more_code_info": ""
    }
  ]
}