{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Shenyang",
      "state": "N/A",
      "country": "China",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82753",
      "recalling_firm": "Neusoft Medical Systems Co., Ltd.",
      "address_1": "No. 177-1 Cuangxin Road",
      "address_2": "Hunnan District",
      "postal_code": "N/A",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution.",
      "recall_number": "Z-1281-2019",
      "product_description": "NeuViz 16 Essence Multi-slice CT Scanner System    Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray  transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral planes taken at different angles.",
      "product_quantity": "2",
      "reason_for_recall": "Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan,  there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.",
      "recall_initiation_date": "20181022",
      "center_classification_date": "20190521",
      "termination_date": "20220825",
      "report_date": "20190529",
      "code_info": "Series numbers    N16E180037E; N16E180047E"
    }
  ]
}