  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "West Chester",
      "address_1": "1302 Wrights Ln E",
      "reason_for_recall": "For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel.  This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.",
      "address_2": "",
      "product_quantity": "183",
      "code_info": "Part numbers:03.010.048 and 03.010.052 with   Lot Numbers:   4874015 4874021 4927262 4954481 4963192  4874016 4874022 4944750 4957586 4963193  4784017 4899564 4944751 4957587 4972003  4874018 4899565 4954478 4957588 4972004  4874019 4927260 4954479 4957589    4874020 4927261 4954480 4963190    4922533 4922975 4981158      4922970 4923015 4981159     4922971 4977241 4981160     4922972 4977242 4981162     4922973 4981156 5035905     4922974 4981157 5035906",
      "center_classification_date": "20150316",
      "distribution_pattern": "Nationwide Distribution",
      "state": "PA",
      "product_description": "Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the Thumb Screw for the Aiming Arm for Titanium Cannulated Tibial Nails-EX (part number 03.010.052).       The Recon Locking Aiming Arm for Lateral Entry Femoral Recon Nails- EX is used when locking the Femoral Nail-EX.  The Aiming Arm for Titanium (TI) Cannulated Tibial Nails-EX is used when locking the Tibial Nail-EX.",
      "report_date": "20150325",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Synthes, Inc.",
      "recall_number": "Z-1281-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70455",
      "termination_date": "20160826",
      "more_code_info": "",
      "recall_initiation_date": "20150217",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}