      "status": "Terminated",
      "city": "Minnetonka",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64875",
      "recalling_firm": "American Medical Systems, Inc.",
      "address_1": "10700 Bren Rd W",
      "address_2": "",
      "postal_code": "55343-9679",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- Nationwide Distribution including the states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NH, NJ, NM, NY,NC, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VT, VA, WA and WI., and the countries of  AUSTRIA, AUSTRALIA, CANADA, BELGIUM, SWITZERLAND, CZECH REPUBLIC, GERMANY, DENMARK, ESTONIA, SPAIN, FINLAND, FRANCE, GREAT BRITAIN, IRELAND, ITALY, LEBANON, NETHERLANDS, NORWAY, SWEDEN, TURKEY, ARGENTINA, BRAZIL, IVORY COST, COLOMBIA, ECUADOR, INDIA, JAPAN, MEXICO, NEW ZEALAND, PANAMA and VENEZUELA.",
      "recall_number": "Z-1279-2013",
      "product_description": "AMS 800¿ Urinary Control System.      The AMS 800¿ Urinary Control System is used to treat urinary incontinence due to reduced outlet resistance (intrinsic sphincter deficiency) following prostate surgery. The AMS 800¿ Urinary Control System is an implantable, fluid filled, solid silicone elastomer device used to treat stress urinary incontinence. It is designed to restore the natural process of urinary control. The device simulates normal sphincter function by opening and closing the urethra at the control of the patient. The AMS 800 system consists of three inter connected components: a cuff, a pump, and a pressure-regulating balloon. The three components are connected with kink-resistant tubing. The AMS 800 can be implanted at either the bulbous urethra or the bladder neck.",
      "product_quantity": "605",
      "reason_for_recall": "American Medical Systems, Inc. is initiating a recall on two components of the AMS 800¿ Urinary Control System,  Control Pump &  Pressure-Regulating Balloon. These products were released despite not meeting all final product release criteria.",
      "recall_initiation_date": "20130301",
      "center_classification_date": "20130510",
      "termination_date": "20140424",
      "report_date": "20130522",
      "code_info": "Model #       Lot #    720172-01 737875001  720172-01 737875002  720172-01 737875003  720172-01 737875004  72400098 714613001  72400098 714613002  72400098 714613003  72400098 714613004  72400098 714613005  72400098 714613006  72400098 714613007  72400098 714613008  72400098 714613009  72400098 714613010  72400098 714613011  72400098 714613013  72400098 714613014  72400098 714613015  72400098 714631001  72400098 714631002  72400098 714631003  72400098 714631004  72400098 714631005  72400098 714631006  72400098 714631007  72400098 714631008  72400098 714631009  72400098 714631010  72404127 786981017  72400098 714631014  72400098 714631015  72400098 714631016  72400098 714631017  72400098 714631018  72400098 714631019  72400098 714631020  72400098 714644001  72400098 714644002  72400098 714644003  72400098 714644004  72400098 714644005  72400098 714644007  72400098 714644008  72400098 714644009  72400098 714644010  72400098 714644011  72400098 714644012  72400098 714644013  72400098 714644014  72400098 714644015  72400098 714644016  72400098 714644017  72400098 714660001  72400098 714660002  72400098 714660003  72400098 714660004  72400098 714660005  72400098 714660007  72400098 714660008  72400098 714660010  72400098 714660012  72400098 714660013  72400098 714660016  72400098 714660017  72400098 714660018  72400098 714660019  72400098 716228001  72400098 716228002  72400098 716228003  72400098 716228004  72400098 716228005  72400098 716228006  72400098 716228007  72400098 716228008  72400098 716228009  72400098 716228010  72400098 716228011  72400098 716228012  72400098 716228013  72400098 716228014  72400098 716228015  72400098 716228016  72400098 716228017  72400098 717297001  72400098 717297002  72400098 717297003  72400098 717297004  72400098 717297005  72400098 717297006  72400098 717297007  72400098 717297008  72400098 717297009  72400098 717297010  72400098 717297011  72400098 717297012  72400098 717297013  72400098 717297014  72400098 717297015  72400098 717297016  72400098 717297017  72400098 717297018  72400098 717297019  72400098 719372001  72400098 719372002  72400098 719372003  72400098 719372004  72400098 719372005  72400098 719372006  72400098 719372007  72400098 719372008  72400098 719372009  72400098 719372010  72400098 719372011  72400098 719372012  72400098 719372013  72400098 719372014  72400098 719372015  72400098 719372016  72400098 719372017  72400098 719372018  72400098 719372019  72400098 731325001  72400098 731325002  72400098 731325003  72400098 731325005  72400098 731325007  72400098 731325008  72400098 731325009  72400098 731325010  72400098 731325011  72400098 731325012  72400098 731325016  72400098 731325017  72400098 731325019  72400098 731325020  72400098 733301001  72400098 733301002  72400098 733301003  72400098 733301004  72400098 733301005  72400098 733301006  72400098 733301007  72400098 733301008  72400098 733301009  72400098 733301010  72400098 733301011  72400098 733301012  72400098 733301013  72400098 733301014  72400098 733301015  72400098 733301016  72400098 733301017  72400098 733301018  72400098 733995001  72400098 733995002  72400098 733995003  72400098 733995004  72400098 733995005  72400098 733995006  72400098 733995007  72400098 733995008  72400098 733995009  72400098 733995010  72400098 733995011  72400098 733995012  72400098 733995013  72400098 733995014  72400098 733995015  72400098 733995016  72400098 735549001  72400098 735549002  72400098 735549003  72400098 735549004  72400098 735549005  72400098 735549006  72400098 735549007  72400098 735549008  72400098 735549009  72400098 735549010  72400098 735549011  72400098 735549012  72400098 735549013  72400098 735549014  72400098 735549015  72400098 735549016  72400098 735551001  72400098 735551002  72400098 735551003  72400098 735551004  72400098 735551005  72400098 735551006  72400098 735551007  72400098 735551008  72400098 735551009  72400098 735551010  72400098 735551011  72400098 735551012  72400098 735551013  72400098 735551014  72400098 735551015  72400098 735551016  72400098 736848001  72400098 736848002  72400098 736848003  72400098 736848004  72400098 736848005  72400098 736848006  72400098 736848007  72400098 736848009  72400098 736848010  72400098 736848011  72400098 736848012  72400098 736848013  72400098 736848014  72400098 736848015  72400098 736848016  72400098 736848017  72400098 736848018  72400098 736852001  72400098 736852002  72400098 736852003  72400098 736852004  72400098 736852005  72400098 736852006  72400098 736852007  72400098 736852008  72400098 736852009  72400098 736852010  72400098 736852011  72400098 736852012  72400098 736852013  72400098 736852014  72400098 736852015  72400098 736852016  72400098 736852017  72400098 740109001  72400098 740109002  72400098 740109003  72400098 740109004  72400098 740109005  72400098 740109006  72400098 740109007  72400098 740109008  72400098 740109009  72400098 740109010  72400098 740109011  72400098 740109012  72400098 740109013  72400098 740110001  72400098 740110002  72400098 740110003  72400098 740110004  72400098 740110005  72400098 740110006  72400098 740110007  72400098 740110008  72400098 740110009  72400098 740110010  72400098 740110011  72400098 740110012  72400098 740110013  72400098 740110014  72400098 740110015  72400098 740110016  72400098 740110017  72400098 740111001  72400098 740111002  72400098 740111003  72400098 740111004  72404127 757650014  72404127 759726003  72400098 740111007  72400098 740111008  72404127 762460014  72400098 740111011  72400098 740111012  72400098 740111013  72400098 740111014  72400098 740111015  72400098 740111016  72400098 740111017  72400098 740111018  72400098 740111019  72400098 740111020  72400098 741413001  72400098 741413002  72400098 741413004  72400098 741413005  72400098 741413006  72400098 741413008  72400098 741413009  72400098 741413010  72400098 741413011  72400098 741413012  72400098 741413013  72400098 741413014  72400098 741413015  72400098 741413016  72400098 741413017  72400098 741413018  72400098 741413019  72400098 741415001  72400098 741415002  72400098 741415003  72400098 741415004  72400098 741415005  72400098 741415006  72400098 741415007  72400098 741415008  72400098 741415009  72400098 741415010  72400098 741415011  72400098 741415012  72400098 741415013  72400098 741415014  72400098 741415015  72400098 741415016  72400098 741416001  72400098 741416002  72400098 741416003  72400098 741416004  72400098 741416005  72400098 741416006  72400098 741416007  72400098 741416008  72400098 741416009  72400098 741416010  72400098 741416011  72400098 741416012  72400098 741416013  72400098 741416014  72400098 741416015  72400098 741416016  72400098 741416017  72400098 741416018  72400098 741417001  72400098 741417002  72400098 741417003  72400098 741417004  72400098 741417005  72400098 741417006  72400098 741417007  72400098 741417008  72400098 741417009  72400098 741417010  72400098 741417011  72400098 741417012  72400098 741417013  72400098 741417014  72400098 741417015  72400098 741417016  72400098 741417017  72400098 741417018  72400098 741417019  72400098 741417020  72400098 741418001  72400098 741418002  72400098 741418003  72400098 741418004  72400098 741418005  72400098 741418006  72400098 741418007  72400098 741418008  72400098 741418009  72400098 741418010  72400098 741418011  72400098 741418012  72400098 741418013  72400098 741418014  72400098 741418015  72400098 741418016  72400098 741418017  72400098 744200001  72400098 744200002  72400098 744200003  72400098 744200004  72400098 744200005  72400098 744200006  72400098 744200007  72400098 744200008  72400098 744200009  72400098 744200010  72400098 744200011  72400098 744200012  72400098 744200013  72400098 744200014  72400098 744200015  72400098 744200016  72400098 744200017  72400098 744200018  72400098 744200019  72400098 745411001  72400098 745411002  72400098 745411003  72400098 745411004  72400098 745411005  72400098 745411006  72400098 745411007  72400098 745411008  72400098 745411009  72404127 805583016  72400098 745411012  72400098 745411013  72400098 745411014  72400098 745411015  72400098 745411016  72400098 745411017  72400098 745411018  72400098 745415001  72400098 745415002  72400098 745415003  72400098 745415004  72400098 745415005  72400098 745415006  72400098 745415007  72400098 745415008  72400098 745415009  72400098 745415010  72400098 745415011  72400098 745415012  72400098 745415013  72400098 745415014  72400098 745415015  72400098 745415016  72400098 745415017  72400098 745415018  72400098 756329001  72400098 756329002  72400098 756329003  72400098 756329004  72400098 756329005  72400098 756329006  72400098 756329007  72400098 756329008  72400098 756329009  72400098 756329010  72400098 756329011  72400098 756329012  72400098 756329013  72400098 756329014  72400098 756329015  72400098 756329016  72400098 756329017  72400098 756329018  72400098 756329019  72400098 756329020  72400098 756330001  72400098 756330002  72400098 756330003  72400098 756330004  72400098 756330005  72400098 756330006  72400098 756330007  72400098 756330008  72400098 756330009  72400098 756330010  72400098 756330011  72400098 756330012  72400098 756330013  72400098 756330014  72400098 756330015  72400098 756330016  72400098 756330017  72400098 756330018  72400098 756330019  72400098 756330020  72400098 757658001  72400098 757658002  72400098 757658003  72400098 757658004  72400098 757658005  72400098 757658006  72400098 757658007  72400098 757658008  72400098 757658009  72400098 757658010  72400098 807441010  72400098 757658012  72404127 728832001  72404127 757650016  72400098 757658015  72400098 757658016  72400098 757658017  72400098 757658018  72400098 757658019  72400098 757658020  72400098 758711001  72400098 758711002  72400098 758711003  72400098 758711004  72400098 758711005  72400098 758711006  72400098 758711007  72400098 758711008  72400098 758711009  72400098 758711010  72400098 758711011  72400098 758711012  72400098 758711013  72400098 758711014  72400098 758711015  72400098 758711016  72400098 758711017  72400098 758711018  72400098 758711019  72400098 758711020  72400098 759739001  72400098 759739002  72400098 759739003  72400098 759739004  72400098 759739005  72400098 759739006  72400098 759739007  72400098 759739008  72400098 759739009  72400098 759739010  72400098 759739011  72400098 759739012  72400098 759739013  72400098 759739014  72400098 759739015  72404127 786977001  72400098 759739017  72400098 759739018  72400098 759739019  72400098 759739020  72400098 761119001  72400098 761119003  72400098 761119004  72400098 761119005  72400098 761119006  72400098 761119007  72400098 761119008  72400098 761119009  72400098 761119010  72400098 761119011  72400098 761119012  72400098 761119013  72400098 761119014  72400098 761119015  72400098 761119016  72400098 761119017  72400098 761119018  72400098 761119019  72400098 761119020  72400098 763923001  72400098 763923002  72400098 763923004  72400098 763923006  72400098 763923007  72400098 763923008  72400098 763923009  72400098 763923010  72404127 738143014  72400098 763923014  72400098 763923015  72400098 763923016  72400098 763923017  72404127 747461005  72400098 765207001  72400098 765207002  72400098 765207003  72400098 765207004  72400098 765207005  72400098 765207006  72400098 765207007  72400098 765207008  72400098 765207009  72400098 765207010  72400098 765207011  72400098 765207012  72400098 765207014  72400098 765207015  72400098 765207016  72404127 747461013  72400098 765207018  72400098 765207019  72400098 765207020  72400098 775889001  72400098 775889002  72400098 775889003  72400098 775889004  72400098 775889005  72400098 775889006  72400098 775889007  72400098 775889008  72400098 775889009  72400098 775889010  72400098 775889011  72400098 775889012  72400098 775889013  72400098 775889014  72400098 775889015  72400098 786992001  72400098 786992002  72404127 752963015  72400098 786992004  72400098 786992005  72400098 786992006  72400098 786992007  72404127 757652011  72400098 786992009  72400098 786992011  72400098 786992013  72400098 786992014  72400098 786992015  72400098 786992016  72400098 790170001  72400098 790170002  72400098 790170003  72400098 790170004  72400098 790170005  72400098 790170006  72400098 790170007  72404127 757652016  72400098 790170009  72404127 757652018  72404127 759726009  72404127 779376002  72404127 780792019  72404127 782798003  72404127 805583008  72400098 790170016  72400098 790170017  72404127 721081001  72404127 721081002  72404127 721081003  72404127 721081004  72404127 721081005  72404127 721081006  72404127 721081007  72404127 721081008  72404127 721081009  72404127 721081010  72404127 721081011  72404127 721081012  72404127 721081015  72404127 721081016  72404127 721081013  72404127 721082001  72404127 721082002  72404127 721081014  72404127 721082004  72404127 721082005  72404127 721082006  72404127 721082007  72404127 721082008  72404127 721082009  72404127 721082010  72404127 721082011  72404127 721082012  72404127 721082013  72404127 721082014  72404127 721082015  72404127 721082016  72404127 721082017  72404127 721082018  72404127 721754001  72404127 721754002  72404127 721754003  72404127 721754004  72404127 721754005  72404127 721754006  72404127 721754007  72404127 721754008  72404127 721754009  72404127 721754010  72404127 721754011  72404127 721754013  72404127 721754014  72404127 721754015  72404127 721754017  72404127 721754018  72404127 721754019  72404127 721754020  72404127 722572001  72404127 722572002  72404127 722572003  72404127 722572004  72404127 722572005  72404127 722572006  72404127 722572007  72404127 722572008  72404127 722572009  72404127 722572010  72404127 722572011  72404127 722572012  72404127 722572013  72404127 722572014  72404127 722572015  72404127 722572016  72404127 722572017  72404127 722572018  72404127 722595001  72404127 722595003  72404127 722595004  72404127 722595005  72404127 722595006  72404127 722595007  72404127 722595008  72404127 722595009  72404127 722595010  72404127 722595011  72404127 722595012  72404127 722595013  72404127 722595014  72404127 722595015  72404127 722595016  72404127 722595017  72404127 722596001  72404127 722596002  72404127 722596003  72404127 722596004  72404127 722596005  72404127 722596006  72404127 722596007  72404127 722596008  72404127 722596009  72404127 722596011  72404127 722596012  72404127 722596013  72404127 722596014  72404127 722596015  72404127 722596016  72404127 722596017  72404127 722596018  72404127 722596019  72404127 723409001  72404127 723409002  72404127 723409003  72404127 723409004  72404127 723409005  72404127 723409006  72404127 723409007  72404127 723409008  72404127 723409009  72404127 723409010  72404127 723409011  72404127 723409012  72404127 723409013  72404127 723409014  72404127 723409015  72404127 723409016  72404127 723409017  72404127 723409018  72404127 723409019  72404127 723414001  72404127 723414003  72404127 723414004  72404127 723414005  72404127 723414006  72404127 723414007  72404127 723414008  72404127 723414009  72404127 723414010  72404127 723414011  72404127 723414012  72404127 723414013  72404127 723414014  72404127 723414015  72404127 723414016  72404127 723414017  72404127 723414018  72404127 723414019  72404127 723414020  72404127 724580001  72404127 724580002  72404127 724580003  72404127 724580004  72404127 724580005  72404127 724580006  72404127 724580007  72404127 724580008  72404127 721081017  72404127 724580010  72404127 746349008  72404127 724580012  72404127 724580013  72404127 724580014  72404127 724580015  72404127 724580016  72404127 724580017  72404127 724580018  72404127 724580019  72404127 725192001  72404127 725192002  72404127 725192003  72404127 725192004  72404127 725192005  72404127 725192006  72404127 725192007  72404127 725192008  72404127 725192009  72404127 725192010  72404127 725192011  72404127 725192012  72404127 725192013  72404127 725192014  72404127 725192015  72404127 725192016  72404127 725192017  72404127 725192018  72404127 725192019  72404127 725204001  72404127 725204002  72404127 725204003  72404127 725204004  72404127 725204005  72404127 725204006  72404127 725204007  72404127 725204008  72404127 725204009  72404127 725204010  72404127 725204011  72404127 725204012  72404127 725204013  72404127 725204014  72404127 725204015  72404127 725204016  72404127 725204017  72404127 725204018  72404127 725204019  72404127 725207001  72404127 725207002  72404127 725207003  72404127 725207004  72404127 725207005  72404127 725207006  72404127 725207007  72404127 725207008  72404127 725207009  72404127 725207010  72404127 725207011  72404127 725207012  72404127 725207013  72404127 725207014  72404127 725207015  72404127 725207016  72404127 725207017  72404127 725207018  72404127 725207019  72404127 725208001  72404127 725208002  72404127 725208003  72404127 725208004  72404127 725208005  72404127 725208006  72404127 725208007  72404127 780792007  72404127 725208009  72404127 725208010  72404127 725208011  72404127 725208012  72404127 725208013  72404127 725208014  72404127 725208015  72404127 725208016  72404127 727984001  72404127 727984002  72404127 727984003  72404127 727984004  72404127 727984005  72404127 727984006  72404127 727984007  72404127 727984008  72404127 727984009  72404127 727984010  72404127 727984011  72404127 727984012  72404127 727984013  72404127 727984014  72404127 727984015  72404127 727984016  72404127 727990001  72404127 727990002  72404127 727990003  72404127 727990004  72404127 727990005  72404127 786977010  72404127 727990007  72404127 727990008  72404127 727990009  72404127 727990010  72404127 727990011  72404127 727990012  72404127 727990013  72404127 727990014  72404127 727990015  72404127 727990016  72404127 727990017  72404127 727990018  72404127 727990019  72404127 727990020  72404127 728830001  72404127 728830002  72404127 728830003  72404127 728830004  72404127 728830005  72404127 728830006  72404127 728830007  72404127 728830008  72404127 728830009  72404127 728830010  72404127 728830011  72404127 728830012  72404127 728830013  72404127 728830015  72404127 728830016  72404127 728830017  72404127 728830018  72404127 728830019  72404127 728832002  72404127 728832003  72404127 728832004  72404127 728832005  72404127 728832006  72404127 728832007  72404127 728832008  72404127 728832009  72404127 728832010  72404127 728832011  72404127 728832012  72404127 728832013  72404127 728832014  72404127 728832015  72404127 728832016  72404127 728832017  72404127 728832018  72404127 728832019  72404127 728837001  72404127 728837002  72404127 728837003  72404127 728837004  72404127 728837005  72404127 728837006  72404127 728837007  72404127 728837008  72404127 728837009  72404127 728837010  72404127 728837011  72404127 728837012  72404127 728837013  72404127 728837014  72404127 728837015  72404127 728837016  72404127 728837017  72404127 728837018  72404127 728837019  72404127 729961001  72404127 729961002  72404127 729961003  72404127 729961004  72404127 729961005  72404127 729961006  72404127 729961007  72404127 729961008  72404127 729961009  72404127 721082003  72404127 729961010  72404127 729961012  72404127 729961013  72404127 729961011  72404127 729961015  72404127 729961016  72404127 729961017  72404127 729961018  72404127 729963001  72404127 729963002  72404127 729963003  72404127 729963004  72404127 729963005  72404127 729963006  72404127 752961004  72404127 729963008  72404127 729963009  72404127 729963010  72404127 729963011  72404127 729963012  72404127 729963013  72404127 729963014  72404127 729963015  72404127 729963016  72404127 729963017  72404127 729963018  72404127 729963019  72404127 731275001  72404127 731275002  72404127 731275003  72404127 731275004  72404127 731275005  72404127 731275006  72404127 731275007  72404127 731275008  72404127 731275009  72404127 731275010  72404127 731275011  72404127 731275012  72404127 731275013  72404127 731275014  72404127 731275015  72404127 731275016  72404127 738143001  72404127 738143002  72404127 738143003  72404127 738143004  72404127 780787012  72404127 738143006  72404127 738143007  72404127 738143008  72404127 738143009  72404127 738143010  72404127 738143011  72404127 738143012  72404127 738143013  72404127 786982012  72404127 738143015  72404127 738143016  72404127 738143017  72404127 738143018  72404127 743171002  72404127 743171003  72404127 743171004  72404127 743171005  72404127 743171006  72404127 743171007  72404127 743171008  72404127 743171009  72404127 743171010  72404127 743171011  72404127 743171012  72404127 744195001  72404127 744195002  72404127 744195003  72404127 744195004  72404127 744195005  72404127 744195006  72404127 744195007  72404127 744195008  72404127 744195009  72404127 744195010  72404127 744195011  72404127 744195012  72404127 744195013  72404127 744195014  72404127 744195015  72404127 744195016  72404127 744195017  72404127 744195018  72404127 744195019  72404127 744195020  72404127 746292001  72404127 746292002  72404127 746292003  72404127 746292004  72404127 746292005  72404127 746292006  72404127 746292007  72404127 746292008  72404127 746292009  72404127 746292010  72404127 746292011  72404127 746292012  72404127 746292013  72404127 746292014  72404127 746292015  72404127 746292016  72404127 746292017  72404127 746292018  72404127 746292019  72404127 746292020  72404127 746349001  72404127 767450015  72404127 746349003  72404127 746349004  72404127 746349005  72404127 746349006  72404127 746349007  72404127 746349009  72404127 746349010  72404127 746349011  72404127 746349012  72404127 746349013  72404127 746349014  72404127 746349015  72404127 746349016  72404127 746349017  72404127 746349018  72404127 746349019  72404127 746349020  72404127 747461001  72404127 747461002  72404127 747461003  72404127 747461004  72404127 738143005  72404127 747461006  72404127 747461007  72404127 747461008  72404127 747461009  72404127 747461010  72404127 747461011  72404127 747461012  72404127 746349002  72404127 747461014  72404127 747461015  72404127 747461016  72404127 747461017  72404127 747461018  72404127 752264001  72404127 752264002  72404127 752264003  72404127 752264004  72404127 752264005  72404127 752264006  72404127 752264007  72404127 752264008  72404127 752264009  72404127 752264010  72404127 752264011  72404127 752264012  72404127 752264013  72404127 752264014  72404127 752264015  72404127 752264016  72404127 752264017  72404127 752264018  72404127 752264019  72404127 752264020  72404127 752961001  72404127 752961002  72404127 752961003  72404127 752963008  72404127 752961005  72404127 752961006  72404127 752961007  72404127 752961008  72404127 752961009  72404127 752961010  72404127 752961011  72404127 752961012  72404127 752961013  72404127 752961014  72404127 752961015  72404127 752961016  72404127 752961017  72404127 752961018  72404127 752963001  72404127 752963002  72404127 752963003  72404127 752963004  72404127 752963005  72404127 752963006  72404127 752963007  72404127 756321003  72404127 752963009  72404127 752963010  72404127 752963011  72404127 752963012  72404127 752963013  72404127 752963014  72404127 759728004  72404127 752963016  72404127 752963017  72404127 752963018  72404127 754367001  72404127 754367002  72404127 754367003  72404127 754367004  72404127 754367005  72404127 754367006  72404127 754367007  72404127 754367008  72404127 754367009  72404127 754367010  72404127 754367011  72404127 754367012  72404127 754367013  72404127 754367014  72404127 754367015  72404127 754367016  72404127 754367017  72404127 754367018  72404127 754367019  72404127 754369001  72404127 754369002  72404127 754369003  72404127 754369004  72404127 754369005  72404127 754369006  72404127 754369007  72404127 754369008  72404127 754369009  72404127 754369010  72404127 754369011  72404127 754369012  72404127 754369013  72404127 754369014  72404127 754369015  72404127 754369016  72404127 754369017  72404127 754369018  72404127 756319001  72404127 756319002  72404127 756319003  72404127 756319004  72404127 756319005  72404127 756319006  72404127 756319007  72404127 756319008  72404127 756319010  72404127 756319011  72404127 756319012  72404127 756319013  72404127 756319014  72404127 759728005  72404127 756319016  72404127 756319017  72404127 756319018  72404127 756321001  72404127 756321002  72404127 762460001  72404127 756321004  72404127 756321005  72404127 756321006  72404127 756321007  72404127 756321008  72404127 756321009  72404127 756321010  72404127 756321011  72404127 756321012  72404127 756321013  72404127 756321014  72404127 756321015  72404127 756321016  72404127 756321017  72404127 756321018  72404127 756321019  72404127 756323001  72404127 756323002  72404127 756323003  72404127 756323004  72404127 756323005  72404127 756323006  72404127 756323007  72404127 756323008  72404127 756323009  72404127 756323010  72404127 756323011  72404127 756323012  72404127 756323013  72404127 756323014  72404127 756323015  72404127 756323016  72404127 756323017  72404127 756323018  72404127 756323019  72404127 756323020  72404127 756324001  72404127 756324002  72404127 756324003  72404127 756324004  72404127 756324005  72404127 756324006  72404127 756324007  72404127 756324008  72404127 756324009  72404127 756324010  72404127 756324011  72404127 756324012  72404127 756324013  72404127 756324014  72404127 756324015  72404127 756324016  72404127 757650001  72404127 757650002  72404127 757650003  72404127 757650004  72404127 757650005  72404127 757650006  72404127 757650007  72404127 757650008  72404127 757650009  72404127 757650010  72404127 757650011  72404127 757650012  72404127 757650013  72404127 786982004  72404127 757650015  72404127 800328012  72404127 757650017  72404127 757650018  72404127 757650019  72404127 757651001  72404127 757651002  72404127 757651003  72404127 757651004  72404127 757651005  72404127 757651006  72404127 757651007  72404127 757651008  72404127 757651009  72404127 757651010  72404127 757651011  72404127 757651012  72404127 757651013  72404127 757651014  72404127 757651015  72404127 757651016  72404127 757651017  72404127 757651019  72404127 757651020  72404127 757652001  72404127 757652002  72404127 757652003  72404127 757652004  72404127 757652005  72404127 757652006  72404127 757652007  72404127 757652008  72404127 757652009  72404127 757652010  72404127 805583010  72404127 757652012  72404127 805583011  72404127 757652014  72404127 757652015  72404127 805583013  72404127 757652017  72404127 757652019  72404127 757654001  72404127 757654002  72404127 757654003  72404127 757654004  72404127 757654005  72404127 757654006  72404127 757654007  72404127 757654008  72404127 757654009  72404127 757654010  72404127 757654011  72404127 757654012  72404127 757654013  72404127 757654014  72404127 757654015  72404127 757654016  72404127 757654017  72404127 757654018  72404127 757654019  72404127 757654020  72404127 758706001  72404127 758706002  72404127 758706003  72404127 758706004  72404127 758706005  72404127 758706006  72404127 758706007  72404127 758706008  72404127 758706010  72404127 758706011  72404127 758706012  72404127 758706013  72404127 758706014  72404127 758706015  72404127 758706016  72404127 758706017  72404127 758706018  72404127 758706019  72404127 758706020  72404127 759724001  72404127 759724002  72404127 759724003  72404127 759724004  72404127 759724005  72404127 759724006  72404127 759724007  72404127 759724008  72404127 759724009  72404127 759724010  72404127 759724011  72404127 759724012  72404127 759724013  72404127 759724014  72404127 759724015  72404127 759724016  72404127 759724017  72404127 759724018  72404127 759724019  72404127 759724020  72404127 759726001  72404127 759726002  72404127 759726004  72404127 759726005  72404127 759726006  72404127 759726007  72404127 759726008  72404127 759726010  72404127 759726011  72404127 759726012  72404127 759726013  72404127 759726014  72404127 759726015  72404127 759726016  72404127 759726017  72404127 759726018  72404127 759726019  72404127 759726020  72404127 759728001  72404127 759728002  72404127 759728003  72404127 759728006  72404127 759728007  72404127 759728008  72404127 759728009  72404127 759728010  72404127 759728011  72404127 759728012  72404127 759728013  72404127 759728014  72404127 759728015  72404127 759728016  72404127 759728017  72404127 759728018  72404127 759728019  72404127 762460002  72404127 762460003  72404127 762460004  72404127 762460005  72404127 762460006  72404127 762460007  72404127 762460008  72404127 762460009  72404127 762460010  72404127 762460011  72404127 762460012  72404127 762460013  72404127 743171001  72404127 762460015  72404127 762460016  72404127 762460017  72404127 786981006  72404127 762460019  72404127 765202001  72404127 765202002  72404127 765202003  72404127 765202004  72404127 765202005  72404127 765202006  72404127 765202007  72404127 765202008  72404127 765202009  72404127 765202010  72404127 765202011  72404127 765202012  72404127 765202013  72404127 765202014  72404127 765202015  72404127 765202016  72404127 765202017  72404127 765202018  72404127 765202019  72404127 767450001  72404127 767450002  72404127 767450003  72404127 767450004  72404127 767450005  72404127 767450006  72404127 767450007  72404127 767450008  72404127 767450009  72404127 767450010  72404127 767450011  72404127 767450012  72404127 767450013  72404127 767450014  72404127 767450016  72404127 767450017  72404127 767450018  72404127 767450019  72404127 778479001  72404127 778479002  72404127 778479003  72404127 778479004  72404127 778479005  72404127 778479006  72404127 778479007  72404127 778479008  72404127 778479009  72404127 778479010  72404127 778479011  72404127 779376001  72404127 780787001  72404127 779376003  72404127 779376004  72404127 779376005  72404127 779376006  72404127 779376007  72404127 779376008  72404127 779376009  72404127 779376010  72404127 779376011  72404127 779376012  72404127 779376013  72404127 779376014  72404127 780786002  72404127 780786003  72404127 780786004  72404127 780787002  72404127 780787003  72404127 780787004  72404127 780787005  72404127 780787006  72404127 780787007  72404127 780787008  72404127 780787009  72404127 780787010  72404127 780787011  72400098 761119002  72404127 780792001  72404127 780792002  72404127 780792003  72404127 780792004  72404127 780792005  72404127 780792006  72404127 780792008  72404127 780792009  72404127 780792010  72404127 780792011  72404127 780792012  72404127 780792013  72404127 780792014  72404127 780792015  72404127 780792016  72404127 780792017  72404127 780792018  72404127 724580011  72404127 782798001  72404127 782798002  72404127 727990006  72404127 782798004  72404127 782798005  72404127 782798006  72404127 782798009  72404127 782798010  72404127 782798011  72404127 782798012  72404127 784784001  72404127 784784002  72404127 784784003  72404127 784784004  72404127 729961014  72404127 784784007  72404127 784784008  72404127 784784009  72404127 784784010  72404127 784784011  72404127 784784012  72404127 784784013  72404127 784784014  72404127 784784015  72404127 784784016  72404127 756319015  72404127 786977002  72404127 786977003  72404127 786977004  72404127 786977005  72404127 786977006  72404127 786977007  72404127 786977008  72404127 786977009  72404127 758706009  72404127 786977011  72404127 786977012  72404127 786977013  72404127 786977014  72404127 786977015  72404127 786977016  72404127 786977017  72404127 786977018  72404127 786981001  72404127 786981002  72404127 786981003  72404127 786981004  72404127 786981005  72404127 786981007  72404127 786981008  72404127 786981009  72404127 786981010  72404127 786981011  72404127 786981012  72404127 786981013  72404127 786981014  72404127 786981015  72404127 786981016  72400098 759739016  72404127 786981018  72404127 786981019  72404127 786981020  72404127 786982001  72404127 786982002  72404127 786982003  72404127 786982005  72404127 786982006  72404127 786982007  72404127 786982008  72404127 786982009  72404127 786982010  72400098 763923012  72404127 786982013  72404127 786982014  72404127 786982015  72404127 786982016  72404127 786982017  72404127 800328001  72404127 800328002  72404127 800328003  72404127 800328004  72404127 800328005  72404127 800328006  72404127 800328007  72404127 800328008  72404127 800328009  72404127 800328010  72404127 800328011  72400098 763923013  72404127 801358001  72404127 801358002  72404127 801358003  72404127 801358004  72404127 801358005  72404127 8013",
      "more_code_info": "58007  72404127 801358008  72404127 801358009  72404127 801358011  72404127 801358012  72404127 801358013  72404127 801358014  72404127 801358015  72404127 801358016  72404127 801358017  72404127 801358018  72404127 801358019  72404127 802432004  72404127 802432016  72404127 802436003  72404127 802436004  72404127 802436005  72404127 802436007  72404127 802436012  72404127 802436014  72404127 802436017  72404127 802436019  72404127 802436020  72404127 805583001  72404127 805583002  72404127 805583004  72404127 805583005  72404127 805583006  72400098 763923018  72404127 805583012  72400098 714631012  72400098 714631013  72400098 740111005  72400098 740111006  72400098 740111009  72400098 745411010  72400098 745411011  72400098 765207017  72400098 786992003  72404127 729963007  72404127 784784005  72404127 724580009  72404127 757652013  72404127 762460018  72400098 757658011  72400098 757658013  72400098 757658014  72400098 786992008  72400098 790170008  72400098 790170010  72400098 790170011  72400098 790170012  72400098 790170013  72400098 790170014  72400098 790170015"
    }
  ]
}