  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Alpharetta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79479",
      "recalling_firm": "Halyard Health, Inc",
      "address_1": "5405 Windward Pkwy",
      "address_2": "N/A",
      "postal_code": "30004-3894",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to states of: AR, CA, CO, GA, IA, LA, MI, MN, NM, TX; and country of Germany.      Distributed internationally to Germany.",
      "recall_number": "Z-1277-2018",
      "product_description": "CORTRAK* 2 Nasogastric/Nasointestinal (NG/NI) Feeding Tubes with Electromagnetic Transmitting Stylet and CORFLO* Enteral Feeding tube, ANTI-IV* Connector, 10 Fr, 140 cm    The device is used in patients that require intermittent or continuous tube feedings via the nasogastric or nasoenteric route.",
      "product_quantity": "185 cases (10 units per case)",
      "reason_for_recall": "The transmitting stylet within CORTRAK* 2 nasogastric/Nasointestinal (NG/NI) Feeding Tubes with ANTI-IV* connector (Product Code 20-9551Trak2A) may be visible within the window of the exit port of the feeding tube.  Based on specifications, the transmitting stylet in the NG tube should not be visible within the port window.",
      "recall_initiation_date": "20171213",
      "center_classification_date": "20180330",
      "termination_date": "20211104",
      "report_date": "20180411",
      "code_info": "Product code 20-9551TRAK2A, Lot No. 7909, 7839, 7876."
    }
  ]
}