  "results": [
    {
      "status": "Terminated",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90218",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the state of KY.",
      "recall_number": "Z-1274-2022",
      "product_description": "Colonovideoscope   Model CF-H180AL",
      "product_quantity": "1 unit",
      "reason_for_recall": "A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding",
      "recall_initiation_date": "20220422",
      "center_classification_date": "20220621",
      "termination_date": "20240313",
      "report_date": "20220629",
      "code_info": "UDI-DI: 04953170339615 Serial number: 2501131",
      "more_code_info": ""
    }
  ]
}