  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67556",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "51 Valley Stream Pkwy",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA Nationwide Distribution in the states of WI, KY, OH, NJ, NC, FL, PA, MA, IA, TX, NY, IL, MI, AZ, CA, and LA.",
      "recall_number": "Z-1272-2014",
      "product_description": "AXIOM Artis Systems    Product usage:  Artis zee is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.",
      "product_quantity": "37",
      "reason_for_recall": "Certain units of the AXIOM Artis Systems are using a flat detector cooling unit with inappropriate connectors that may possibly leak, causing a system failure and possible drop out of the gantry system.",
      "recall_initiation_date": "20140207",
      "center_classification_date": "20140321",
      "termination_date": "20141230",
      "report_date": "20140402",
      "code_info": "Model nos. 7412807, 7727717, 7728392 with serial numbers: 35020, 34999,35098  35113  35042  35061  35064  35073  35014  35069  35053  35055  35056  35054  35008  35057  35044  35018  35097  35108  35046  35048  35050  35045  35004  35019  46004  46005  40000  40009  40011  40006  40007  40010  44016  44005  44011"
    }
  ]
}